Items Required - ABB Navigator 550 Manual

Hydrazine and sodium wet-sections, flowmeter upgrade procedures single-stream and multi-stream
Hide thumbs Also See for Navigator 550:

Advertisement

3 Safety
Warning.
These procedures must be carried out by a trained
technician.
CHEMICAL
Ensure personal protective equipment (PPE) such as
and
gloves
eye protection
maintenance.
Observe all health and safety procedures for handling
chemicals.
To familiarize yourself with handling precautions, dangers
and emergency procedures, always review the Material
Safety Data Sheets prior to handling containers,
reservoirs and delivery systems that contain chemical
reagents and standards.
Take care if cleaning any spillages and observe all relevant
safety instructions. Wipe up any spillages using clean
water.
Perform general cleaning of the wet-section using a damp
cloth only – mild detergent can be used as a cleaning aid.
Do not use Acetone or any organic solvents.
ELECTRICAL
Isolate all high voltage supplies to the transmitter before
performing bottle / label replacement procedures.
The wet-section is vulnerable to electrostatic damage.
Wear an anti-static strap or dismantle the wet-section on
an anti-static workbench.
Ensure all electrical connections are kept dry at all times.
GENERAL
Shut off the external sample supply to the wet-section
before performing replacement procedures. When a
procedure is complete, restore power to the transmitter
and sample to the wet-section at the correct flow rate. If
necessary, calibrate the wet-section – refer to the
wet-section
Operating
OI/ASO550-EN
or
calibration instructions.
Dispose of the old components in accordance with the
guidelines contained in the Operating instructions
(sodium OI/ASO550-EN or hydrazine OI/AHM550-EN).
2
INF13/141–EN Rev. B | Navigator 550 | Hydrazine and sodium wet-sections | Flowmeter upgrade procedures (single-stream and multi-stream)
are worn during any
Instructions
(sodium
hydrazine
OI/AHM550-EN)

4 Items required

4.1 Single-stream wet-sections (hydrazine / sodium)
1 x flowmeter upgrade kit, part number AW500 040
4.2 Multi-stream wet-sections (sodium)
3 x flowmeter upgrade kits, part number AW500 041
Fig. 4.1 Flowmeter assembly upgrade – flowmeter location(s)
(sodium wet-section shown)
for
Flowmeter assembly
and flowmeter cable(s)
3 Safety