Grundfos BACMON Installation And Operating Instructions Manual page 8

Real-time bacteria monitoring
Table of Contents

Advertisement

7.5.1 "Last result"
Home
Status
Sample ID: 29630
Analysis time: 2016-04-25 16:09:54
Bacteria concentration: 57430 /ml
Non-bacteria concentration: 98338 /ml
Flow cell: 91%
Measured volume: 6.13095 μl
Bad volume: 0.0192907 μl
Fig. 12 "Last result"
This submenu shows the data of the latest sample taken.
7.5.2 "View sensors"
Home
Status
Board temperature: 30 °C
Flow cell temperature: 20.59 °C
Flow cell pressure: 2.003 bar
Fig. 13 "View sensors"
This submenu shows the values of the different measurements.
7.5.3 "Current alarms"
Home
Status
2016-04-25 16:20:07 E41
Details about selected alarm:
Flow cell fouling is too high
Fig. 14 "Current alarms"
This submenu shows alarms related to the hardware. It shows an
E-number which can be looked up in section
warnings. The product will stop measuring when hardware alarms
are active.
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
Clearing alarms
1. Highlight the alarm you wish to clear and press [OK].
2. Highlight "Yes", press [OK] to clear the alarm or press [Right]
to choose "No" and [OK] to abort.
8
Service
Settings
Service
Settings
Service
Settings
8.2 Alarms and
7.5.4 "Current warnings"
Home
Status
2016-04-25 16:20:07 E41
Details about selected warning:
Flow cell fouling is too high
Fig. 15 "Current warnings"
This submenu shows warnings related to the hardware. It shows
an E-number which can be looked up in section
warnings.
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
7.5.5 "Current water quality alarms"
Home
Status
Bacteria concentration
Bad volume
Details about selected alarm:
Fig. 16 "Current water quality alarms"
This submenu shows alarms related to the water quality. The
product will continue taking samples when water quality alarms
are active.
Some alarms may not be described in this manual, but they can
help Grundfos service specialists find the error in the product, if
needed.
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
Clearing alarms
1. Highlight the alarm you wish to clear and press [OK].
2. Highlight "Yes", press [OK] to clear the alarm or press [Right]
to choose "No" and [OK] to abort.
Service
Settings
8.2 Alarms and
Service
Settings