Table 3 – Proximity fields from RF wireless communications equipment
Test Frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Phenomenon
Electrical fast transients/burst
Surges line-to-line
Conducted disturances
induced by RF fields
Voltage dips
Voltage interruptions
Band
(MHz)
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
Table 4 – Input a.c. power Port
Basic EMC Standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
11
Immunity test levels
Professional healthcare
facility environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
Immunity test levels
Professional healthcare
facility environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz
and 80MHz
80%AM at 1kHz
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º
and 315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
0% UT; 250/300 cycles
Need help?