User Manual R I W I T N E S S ™ Embryology Heated Plate +45 46 79 02 02 | customerservice@origio.com | coopersurgical.com Only Research Instruments Ltd, Bickland Industrial Park, Falmouth, Cornwall TR11 4TA, UK 0120 Document 6-70-807UM(6) | DRF 4740 | 29 April 2019...
Contents Research Instruments Ltd CONTENTS SECTION 1 - PREFACE SECTION 2 - INTRODUCTION TO RI WITNESS Indications for Use for RI Witness Embryology Heated Plate Contraindications Applicable Part Numbers Related Documents Compatibility Installation SECTION 3 - SAFETY WARNINGS Glossary of Safety/Information Symbols...
Page 3
Contents CONTENTS Changing the Temperature Setpoint Using the Device Changing the Temperature Setpoint Using a PC and RI Witness WorkArea Software Temperature Calibration ITO Glass Window Calibration Using Built-In User Interface Full 5-channel Calibration Using PC and RI Witness WorkArea Software...
The latest version of the User Manual can be downloaded from Availability of RI Witness for clinical use is dependent on the regulatory approval status of RI Witness in software.research-instruments.com. The RI Witness manual belongs with the RI Witness system and the country where the device is sold.
The system should be operated by qualified and trained personnel only. DO NOT disassemble or modify any part of the RI Witness Embryology Heated plate, or substitute any component for any other. Doing so may result in damage to samples. This voids the warranty and/or service contract.
Section 3 Section 3 Safety Warnings Research Instruments Ltd Research Instruments Ltd Safety Warnings Glossary of Safety/Information Symbols Source: ISO 15223-1 Symbol Meaning BS EN 60601-1 Symbol Meaning This way up Do not dispose of product with normal waste Disposal of according to EU WEEE Directive Fragile, handle with care In accordance with Annex II of the European Medical Device Directive 93/42/EEC, as amended...
RFID Reader Environment Guidance and Manufacturer’s Declaration (IEC 60601-1-2) An RI Witness system uses Radio Frequency Identification (RFID) readers to monitor a work area. — Electromagnetic Emissions Readers detect RFID tagged containers that are placed in the work area.
40 % UT 40 % UT hospital environment. (60 % dip in UT) for 5 (60 % dip in UT) for 5 If the user of RI Witness IEC 61000-4-11 cycles cycles requires continued 70 % UT...
RI Witness Embryology Heated Plate Specification Table SECTION 4 - PRODUCT OVERVIEW RI Witness is a system which operates within an assisted reproduction (AR) clinic setting and provides a method of identifying human samples throughout an AR cycle (from egg and sperm collection to...
Once the Windows operating system has recognised the devices within the Embryology Heated Plate, open the RI Witness WorkArea software. To verify that the RI Witness Work Area software Place only plastic or glass ware on the window heated area.
The temperature setpoint is applied to all heating channels and is set using the following procedure, or Place the probe of a calibrated thermometer in good thermal contact with the surface. from within the RI Witness WorkArea software. Refer to the RI Witness Software Manual (6-70-121UM) for further information.
Tube Reader Antenna Connections 1. Open the RI Witness WorkArea software and navigate to the WorkArea Status window. 2. Click on the yellow triangle or the (i) icon then click WorkArea Settings, then Connected Devices, then Temperature Controller, then Check Calibration. The screen will now show the current temperature and calibration offsets of the five different heated areas.
When multiple alarms are active, the highest priority alarm will be shown or the icon in the RI Witness WorkArea Software, and on the device the Status Indicator Light.
Section 6 Section 6 Troubleshooting Research Instruments Ltd Research Instruments Ltd Troubleshooting Alarm System Testing Alarm Fault Priority Fault Description Alarm Solution Code Condition Actions In order to test functionality of the alarm system hold down the button for 3 seconds when there ITO Window Low / Heating channel has...
Page 15
Section 6 Section 6 Troubleshooting Research Instruments Ltd Research Instruments Ltd Troubleshooting Alarm Fault Priority Fault Description Alarm Solution Alarm Fault Priority Fault Description Alarm Solution Code Condition Actions Code Condition Actions ITO Window Medium Heating channel has This may be caused ITO Window Medium Heating controller...
If RI determines that the product fails to RI Witness Embryology Heated Plate and Tube Reader Antenna may be cleaned with a soft cloth and conform to that warranty during the one-year period, RI will repair or replace the product, at RI’s mild detergent.
Section 9 Section 9 Returning Product to RI for Repair Research Instruments Ltd Research Instruments Ltd Returning Product to RI for Repair SECTION 9 - RETURNING PRODUCT TO RI FOR REPAIR Customer Service Contact details: Tel: +45 46 79 02 02 Please refer to the ‘Troubleshooting’...