Eaton K1650 Instruction Manual

Eaton K1650 Instruction Manual

Mtl alternator purge gas analysers

Advertisement

Instruction manual
MTL gas analysers & systems
K1650 & K6050
MTL alternator purge gas analysers
January 2017
INM MTL 130-0126 Rev 13

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the K1650 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Eaton K1650

  • Page 1 Instruction manual January 2017 MTL gas analysers & systems INM MTL 130-0126 Rev 13 K1650 & K6050 MTL alternator purge gas analysers...
  • Page 2: Declaration Of Conformity

    DECLARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods. However, you can find a copy of the latest version at - http://www.mtl-inst.com/certificates INM MTL 130-0126 Rev 13...
  • Page 3: Table Of Contents

    K1650 and K6050 alternator purge calibration ........
  • Page 4 THIS PAGE IS LEFT INTENTIONALLY BLANK INM MTL 130-0126 Rev 13...
  • Page 5: Introduction

    Gas analysis takes place using thermal conductivity measurements made using a highly-stable, non-depleting katharometer sensor. The K1650 is supplied with zener barriers to provide an intrinsically safe interface between the sensor unit and the electronics allowing the sensor to be located in a hazardous area.
  • Page 6: Specification

    Stainless Steel System- ” diameter tube on the sample inlet and sample outlet. Inlet and Outlet suitable for 2.11 Analogue outputs K1650 4 to 20mA proportional to 0 to 100% of selected range Maximum load 500 ohms K6050 0 to 1V proportional to 0 to 100% of selected range...
  • Page 7: Alarm Outputs (K1650 Only)

    2.12 Alarm outputs (K1650 only) 4 alarms, volt free C/O contacts rated at 30V ac or dc, 1A - normally energised Alarm 1 – configurable as either concentration of Carbon Dioxide in Air or ‘Fault’ Alarm 2 - configurable as either concentration of Hydrogen in Carbon Dioxide or ‘Hydrogen Purity Range Selected’...
  • Page 8: Installation

    3.3.3 Routing of wires To minimise the pick-up of electrical noise all signal wiring should be routed away from power cables. The K1650 is shipped with a pre-assembled sensor cable. See Section 4.7 for details. INM MTL 130-0126 Rev 13...
  • Page 9: Sampling And Piping

    Sampling and piping The connections are captive seal compression fittings suitable for 0.25” (or 6mm) diameter tube or double ferrule stainless steel compression fittings suitable for ” diameter tube on the sample inlet and sample outlet. NOTE Optimum analyser performance is achieved with the flow rate specified in Section 4.7 Consider also installing pipe work for a calibration gas inlet - this avoids the need to remove the sample connection when connecting a calibration gas.
  • Page 10 Range / Fault Alarm 120-0404-5 Connect to intrinsically safe earth M5 mounting studs (x 6) Cutout required at rear of panel in mounting surface 16mm length 186 x 138mm (centred) Figure 1 - Panel mounting dimensions (K1650) INM MTL 130-0126 Rev 13...
  • Page 11 113.5 163.5 MTL Gas analyser MTL Gas analyser remote sensor remote sensor 120-0308-6 120-0308-6 Figure 2a - Standard sample system Figure 2b - Stainless steel sample system Ø14 K6050 MTL Alternator Purge Monitor View Edit View Edit Meas Meas Hazardous Hazardous Hazardous Hazardous...
  • Page 12: Flow Meter

    K1650 The remote sensor enclosure (see Figure 2) is fitted with a flow meter. Electronic module installation (K1650) The display & control unit is designed to be panel mounted, refer to Figure 1 for further details. Sensor and control/display-unit connections CAUTION The sensor and control unit are matched during manufacture.
  • Page 13 Safe Earth Sensor cable - 16 x 0.2 twisted 30 metres max. REMOTE SENSOR UNIT REAR VIEW OF PANEL Figure 5 - Connections when sensor is in a hazardous area (K1650) Figure 2a - Standard Sample System Connect to Non-hazardous Intrinsically...
  • Page 14: Control/Display Unit Connections

    Figure 7 - Connections at rear of control/display unit Figure 7 shows the connections provided at the rear of the K1650, including the 4/20mA analogue output (terminals 19 & 20), the alarm outputs (terminals 1-12) and the AC power inputs (16-18).
  • Page 15 K1550 model mounted in a hazardous area Connections to this display/control unit are provided on DIN-rail mounted terminals to the left of the display. Terminal identities are provided on a label below them - see Figure 8. M20 threaded holes are provided in the bottom of the flameproof enclosure for these connections - see Figure 3.
  • Page 16: Operation

    NOTE During initialisation the K1650’s 4 to 20mA output goes to 2mA to indicate an invalid measurement. If the instrument is outside of its operating range (fault code displayed) then the 4 to 20mA output goes to 3.7mA to indicate this. The K6050, which has 0 to 1V analogue outputs, gives 0v during initialisation and 0V or 1V for fault conditions.
  • Page 17: Making Measurements

    Making measurements The sample flow rate should be set at 100-300 ml/min. It is essential that the range setting is appropriate to the measurement being made. The ranges are as follows • Range 1: 0 to 100% Carbon Dioxide in Air •...
  • Page 18: Viewing And Changing The K1650 Alarm Settings

    Viewing and changing the K1650 alarm settings Press the “View” button to view the alarm settings. Use the “ ” or “ ” buttons to scroll through the alarm screens. Measurement Mode 98.5% H H in Air View Press CO /Air...
  • Page 19: Fault Codes

    NOTE Hysteresis is the difference between the levels at which an alarm triggers and resets. For example, an instrument with a 100% range with an alarm set to mode = HIGH, level = 5% and hysteresis = 1% will trigger the alarm as the reading exceeds 5%, and reset it once the level has fallen to 4%.
  • Page 20: Calibration

    WARNING The analogue output is set to 2mA (K1650) or 0v (K6050) while the instrument is being calibrated. Ensure that any control loops that are connected to the instrument are disabled prior to verifying or calibrating the instrument.
  • Page 21: Calibration Gases

    Calibration gases Three are required. • 100% Hydrogen • 100% Carbon Dioxide • Dry Air The diagram that follows illustrates the calibration process. 5.3.1 Analysers working at alternator casing pressure When this is the case the analyser must be calibrated at the appropriate pressure(s) for best accuracy.
  • Page 22: K1650 And K6050 Alternator Purge Calibration

    5.3.2 K1650 and K6050 alternator purge calibration Figure 12 - Purge calibration menu options INM MTL 130-0126 Rev 13...
  • Page 23: Servicing

    The katharometer sensor is non-depleting and will last for a long time if it is not subjected to misuse. If a replacement is required, it must be returned to Eaton’s MTL product line, or their agent, because the sensor can only be replaced with the use of specialist equipment.
  • Page 24 MTL Instruments GmbH, Heinrich-Hertz-Str. 12, 50170 Kerpen, Germany UNITED ARAB EMIRATES Cooper Industries/Eaton Corporation Tel: +49 (0)22 73 98 12 - 0 Fax: +49 (0)22 73 98 12 - 2 00 Office 205/206, 2nd Floor SJ Towers, off. Old Airport Road, E-mail: csckerpen@eaton.com...

This manual is also suitable for:

Crouse-hinds seriesK6050

Table of Contents