Download Print this page

Replace Covers - Agilent Technologies E5515BU-863 Installation Note

Wireless communication test set, adds cdma2000, wcdma and 1xev capability to the e5515b , kit

Advertisement

Install Option Label
1. Install the option label, 7121-7884, on the back of the instrument.
a. Don't cover up the old option label if the options on the old label
are not also on the new label.
Remove Non-compatible Applications:
1. Remove all non-compatible Applications from the test set. Removing older
Application revisions prevents improper operation of the test set. The test set
must still be connected to the PC LAN port to use these tools, see Table 1 for
compatibility.
a. Run the Agilent 8960 File Utility
b. Select the Direct Connection (Single test set) button.
c. Select the Test Application tab.
d. Highlight the old revision(s) of Application(s) and select the Delete From
test set button. Do not delete the Applications just loaded.

Replace covers:

1. Replace the top cover and secure it with the 26 screws.
2. Slide the outer shroud (cover) on.
3. Replace the screw on the rear
4. Attach the rear bumpers and handles.
8