Proline Prowirl 200
Endress+Hauser
Terminal assignment
Transmitter
The order code is part of the extended order code. For detailed
information on the features of the device and the structure of the
extended order code →  7.
Connection versions
Order code for
"Output"
Option A
1)
Option B
1)
Option C
1) 2)
Option D
1) 3)
Option E
1) 4)
Option G
1)
Output 1 must always be used; output 2 is optional.
2)
The integrated overvoltage protection is not used with option D: Terminals 5 and 6
(current input) are not protected against overvoltage.
3)
FOUNDATION Fieldbus with integrated reverse polarity protection.
4)
PROFIBUS PA with integrated reverse polarity protection.
Safety-related values
The order code is part of the extended order code. For detailed
information on the features of the device and the structure of the
extended order code →  7.
Terminal numbers
Output 1
1 (+)
2 (-)
3 (+)
4-20 mA HART
(passive)
Pulse/frequency/
4-20 mA HART
(passive)
4-20 mA HART
4-20 mA analog
(passive)
Pulse/frequency/
4-20 mA HART
(passive)
Pulse/frequency/
FOUNDATION
Fieldbus
Pulse/frequency/
PROFIBUS PA
Output 2
4 (-)
5 (+)
-
switch output
(passive)
(passive)
4-20 mA current
switch output
input (passive)
(passive)
switch output
(passive)
switch output
(passive)
XA01148D
Input
6 (-)
-
-
-
-
-
19