Guidance and manufacturer's declaration – electromagnetic immunity
Willis the Whale compressor nebulizer system is compliant for each IMMUNITY test specified by the
standard, e.g. IMMUNITY test level.
Immunity test
IEC 60601-1-2 test level Compliance level
Electrostatic
±8 kV contact
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV,
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
±15 kV air
Radiated RF
10V/m
10V/m
EM fields
80MHz-2.7GHz
80MHz-2.7GHz
80% AM at 1kHz
80% AM at 1kHz
IEC 61000-4-3
Electrical fast
±2 kV
±2 kV
transient/burst
100 kHz repetition
100 kHz repetition
frequency
frequency
IEC 61000-4-4
Surge
±0.5 kV,
±0.5 kV,
±1 kV line-to-line;
±1 kV line-to-line;
IEC 61000-4-5
±0.5 kV, ±1 kV and
±0.5 kV, ±1 kV, and
±2 kV line-to-ground
±2 kV line-to-ground
Conducted
3V
3V
disturbances
0.15MHz-80MHz
0.15MHz-80MHz
induced by
6V in ISM and
6V in ISM and
RF fields
amateur radio bands
amateur radio bands
between 0.15MHz
between 0.15MHz
IEC 61000-4-6
and 80MHz
and 80MHz
80% AM at 1kHz
80% AM at 1kHz
a
Voltage
0% U
: 0.5 cycle
0% U
T
dips, short
At 0°, 45°, 90°, 135°,
At 0°, 45°, 90°, 135°,
interruptions
180°, 225°, 270°,
180°, 225°, 270°,
and voltage
and 315°
and 315°
variations
0% U
: 1 cycle
0% U
T
on power
70% U
: 25/30 cycles
70% U
T
supply input
Single phase at 0°
Single phase at 0°
lines
0% U
:
0% U
T
IEC61000-4-11
250/300 cycles
250/300 cycles
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
12
Electromagnetic environment – guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
a
: 0.5 cycle
Mains power quality should be that
T
of a typical commercial or hospital
environment. If a dips or an interruption
of mains power occurs, the current of
Willis the Whale compressor nebulizer
: 1 cycle
T
system may be dropped off from
: 25/30 cycles
T
normal level, it may be necessary to
use uninterruptible power supply or
:
a battery.
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
7.
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
Band
Service
a
Modulation
Frequency
(MHz)
(MHz)
385
380-390
TETRA 400 Pulse
modulation
18 Hz
c
450
430-470
GMRS 460
FM
FRS 460
±5kHz
deviation
1 kHz sine
710
704-787
LTE-Band 13,
Pulse
745
17
modulation
780
217 Hz
810
800-960
GSM
Pulse
870
800/900,
modulation
930
TETRA 800,
18 Hz
Iden 820,
CDMA 850,
LTE Band 5
1720
1700-1990
GSM 1800;
Pulse
1845
CDMA 1900;
modulation
1970
GSM 1900;
217 Hz
DECT;
LTE Band
1, 3, 4, 25;
UMTS
2450
2400-2570
Bluetooth,
Pulse
WLAN,
modulation
802.11
217 Hz
b/g/n,
RFID 2450,
LTE Band 7
5240
5100-5800
WLAN
Pulse
5500
802.11
modulation
5785
a/n
217 Hz
NOTE:
a. For some services, only the uplink frequencies are included.
b. The carrier shall be modulated using a 50% duty cycle square wave signal.
c. As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does
not represent actual modulation, it would be worst case.
ENGLISH
b
Maximum
Distance
IMMUNITY
Compliance
power
(m)
test level
level
(W)
(V/m)
1.8
0.3
27
27
2
0.3
28
28
0.2
0.3
9
9
2
0.3
28
28
2
0.3
28
28
2
0.3
28
28
0.2
0.3
9
9
13