Specifications - Philips Respironics Wisp Youth Instructions For Use Manual

Nasal mask
Table of Contents

Advertisement

Available languages

Available languages

Specifications

Warning: The technical specifications of the mask are provided for the healthcare professional to determine
if it is compatible with the CPAP or bi-level therapy device. If used outside these specifications, or if used
with incompatible devices, the mask may be uncomfortable, the seal of the mask may not be effective,
optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect device function.
Pressure Flow Curve
60
50
40
30
20
10
0
0
Resistance
Drop in Pressure at
All sizes
Deadspace Volume
P
S/M
L
XL (adult only)
Sound Levels
A-weighted Sound Power Level 27 dBA
A-weighted Sound Pressure Level at 1 m 19 dBA
Disposal
Dispose of in accordance with local regulations.
Storage Conditions
Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing
5
10
Pressure cm H
50 SLPM
0.8 cm H
O
2
37 mL
49 mL
73 mL
85 mL
15
20
O
2
100 SLPM
3.2 cm H
O
2
6
25
30