gbo DUODYNATOR 1049 User Manual

Interference current stimulation device

Advertisement

Quick Links

Interference Current Stimulation Device
®
DUODYNATOR
1049
User Manual
Medizintechnik
 gbo
AG 2008
Version 1.3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DUODYNATOR 1049 and is the answer not in the manual?

Questions and answers

Summary of Contents for gbo DUODYNATOR 1049

  • Page 1 Interference Current Stimulation Device ® DUODYNATOR 1049 User Manual Medizintechnik  gbo AG 2008 Version 1.3...
  • Page 2 DUODYNATOR 1049 The gbo AG has taken care in preparation of this manual, but makes no expressed or implied warranty of any kind and assume no responsibility for errors or omissions. All rights reserved. No part of this manual may be reproduced, in any form or by any means (electronic, mechanical, or otherwise) without the prior written permission of the gbo AG.
  • Page 3: Notations, Warnings And Safety Precautions

    - text appears on the display of the device; Warnings and safety precautions Warning! Warnings which have to be observed by all means! Caution! Observe the instructions for use! Note! Information that will facilitate your work.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 4: Table Of Contents

    Balance regulator Changing treatment parameters 3.4.1 Treatment time 3.4.2 Surge and dynamics parameter Key description 3.5.1 Channel keys 3.5.2 Treatment time key 3.5.3 OK key 3.5.4 ESC key 3.5.5 Menu key 3.5.6 Softkey  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 5 Failures: Application errors: MAINTENANCE Safety Controls Cleaning, Disinfection and Care 8.2.1 Electrode Fleece, Electrode Pockets or Felts WARNINGS AND SAFETY PRECAUTIONS EXPLANATION OF THE SIGNS USED TECHNICAL DATA ACCESSORIES APPENDIX: PROGRAMS INDEX EMC-COMMENTS  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 6: Introduction

    Therapeutic frequencies, especially for gymnastics, muscle training, muscle detoning, lipolysis, fat burning and edema treatment are available. It offers also the possibility of editing programs by the therapist himself.  gbo Medizintechnik AG 2008...
  • Page 7: Operating Concept

    • When turning the intensity regulator without activated channel. Triad gong • At the end of the faultless automatic check at the start of operation. • At the end of therapy. Table 2: Acoustical user support  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 8: View Of Duodynator ® 1049

    1049 Mode keys Channel keys Softkeys Graphic display Softkeys Channel keys Mode keys I + II III + IV DUODYNATOR 1049 Intensity regulator Treatment Modification MENU Intensity regulator channels I + II time key knob channels III + IV Figure 1: Device view...
  • Page 9: Short Instructions: Treatment With Interference Currents

    The intensity can only be increased if there is an active channel. The selection is automatically removed after 15 seconds if there is no user interaction. If you try to increase the intensity without an active channel, the following note will appear: Select channel!  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 10 ® DUODYNATOR 1049 13. At the end of the therapy the triad gong will sound. The current to the patient will be automatically reduced.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 11: Start Of Operation

    5. Take the voltage selector out of the receiver and set the correct mains voltage. Warning! The scotch tape over the mains socket is easy to remove. Connect the current stimulation device with the mains cable to a grounded power outlet.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 12: Basic Settings

    . Select your desired language. Store your choice by pressing the . By pressing twice the OK key you will leave the menu. Now you are ready to start. Further details on possible adjustments are described in chapter 5.1 Modification of the settings.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 13: Description Of Function

    When the device is switched on, the last used therapy programs are loaded. It is possible to start treatment imediately at the point you stopped the last time you used the device: the device always remembers the state  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 14: Galvanic Current

    3.1.5 Galvanic current The DUODYNATOR 1049 offers the possibility to emit galvanic current. This can be done with channel IV only and could be selected as single step program. Galvanic current is used to stimulate the circulation in the area to be treated. It has alleviating characteristics.
  • Page 15 Sport Electromass. 0.0 V 0 mA Shakingmass. Reha Mobilisation Musclepump Trauma 15:00 Menu Lymph.drain. 90-110 Hz Edema Pain 0.0 V 0 mA Universal Single step Figure 4: Current program list on right side.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 16: Setting Output Intensity

    3.4.1 Treatment time Depending on the type of current program the therapy time defined in the basic settings will be suggested. To change the therapy time proceed as follows: 1. Press Treatment time key  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 17: Surge And Dynamics Parameter

    • When dynamic is “on” and the intensity is modified, dynamic is turned off for 5 seconds and then turned on back again.  gbo Medizintechnik AG 2008...
  • Page 18: Key Description

    Figure 8: ESC key 3.5.5 Menu key Pressing the will open up the menu. This can only be done when no MENU key therapy is active. Figure 9: MENU key  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 19: Softkey

    The device remembers the last used programs in MFmod mode and load the channels with these programs. Figure 12: -Key  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 20: Intensity Monitoring

    4. Please observe by all means the notice on the electrode position (see chapter 6.1 Electrode Positioning). We at gbo AG have integrated a special intensity monitoring routine in the device software in order to get these application problems under control. This safety switching avoids an unintended high modification of the intensity of current and switches off the respective channel in such a case.
  • Page 21: The Menu

    Once the device has been switched on and is ready, you can start immediately with the standard settings. To modify the unit settings proceed as follows: 1. Press the . You are in the MENU (see Figure 13: Menu). MENU key  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 22 Volume of beeper 0 – 3 Contrast Display 0 – 8 Language German German English * These languages are not active on delivery. Italian* Spanish* Portuguese* Russian* French* Polish* Table 3: Unit settings  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 23: Loading A User Defined Program

    • If no channel was activated when the menu was opend, the following message appears after pressing the OK key Select channel! • Press one of the to load the program into the desired channel. channel keys  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 24: Editing Of User Defined Programs

    When this step was denoted as empty (indicated by OK key “-“), default parameters will be loaded. Otherwise you will see the details of the selected step:  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 25 All steps will be erased then. The loop parameter can be changed at the program step selection mode. Move the highlight to Loops, confirm with the , and modify its value straight forward. OK key  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 26: Cable Check

    12. Exit the Menu Cable Check by pressing the ESC key 13. On the display the Main menu appears again. 14. Exit the Main menu by pressing the ESC key 15. On the display the therapy display appears again.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 27: Electrodes

    7. Apply the holding strap in a way that the electrode fits completely. 8. The sponge pockets or intermediate layers have to be cleaned after the expiration of treatment, see chapter 8.2.1. Note! The above mentioned applies for the application of suction electrodes accordingly. Warning!  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 28: Single-Pole Electrodes

    Electrode Surface Maximum current Duo electrode small 12 cm 24 mA Duo electrode standard 27 cm 54 mA Table 4: maximum currents in duo electrodes  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 29: Behavior In Case Of Failures

    − This means that the patient circuit has been interrupted. Check the electrodes, cables and connections. ♦ Select channel! Suggestion: − You turned the intensity regulator, but no channel has been activated. Select a channel and try again.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 30: Maintenance

    2. Leave the parts over the night in the neutral soap-suds or tap water. 3. Take them out, and wash them several times thoroughly out under cold or warm running water. 4. Let them drop, but do not wring them out.  gbo Medizintechnik AG 2008...
  • Page 31: Warnings And Safety Precautions

    • The electrodes must adhere completely to the skin in order to avoid excessive local current density. Otherwise there can occur erythemas, and sometimes even burns. • In case of any visible failures please contact immediately gbo AG or one of the service agencies authorized by gbo AG.
  • Page 32: Technical Data

    Pause duration: 2.5 ms - 5 s Frequency: 0.1 - 200 Hz max. output voltage (per channel): 100 V peak gbo AG reserves the right to modify the design and specification without prior notice.  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 33: Accessories

    ® DUODYNATOR 1049 12 Accessories Article Article number with user manual 019-0-1049 DUODYNATOR 1049 NEOSERV 1044 with user manual 019-0-1044 Accessories Distribution cable I (channel I and II) 019-0-0010 Distribution cable II (channel III and IV) 019-0-0011 Patient cable set (1 black and 1 white) red for channel I...
  • Page 34 31 Supply change-over 12 elektrode positioning 28 Surge 14 function keys 8 transport 12 transverse current flow 29 Galvanic current 15 graphic display 8 Warning 12, 15, 28, 31, 32 Warnings 12, 32  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 35: Appendix: Programs

    Treatment time: Fixed to 30 min. Electromassage Endosun 0° Endosun 90° 0.5 – 3 Hz Steps 1 + 2 combined for MFmod 10 seconds 10 seconds therapies. Shakingmassage 3 – 10 Hz Sport 20 Hz only MFmod 5s/5s/20s  gbo AG 2004...
  • Page 36: Emc-Comments

    NEOSERV 1044. If such an order is necessary nevertheless, the unit must be observed in order to check it for the intentional operation. You find more EMC-comments in the chapter “Warnings and Safety Precautions” of this manual as well as in the Technical Information on the next two pages.  gbo AG 2004 Version 1.2...
  • Page 37 IEC 61000-4-8 location in a typical commercial or hospital environment. Note: U is the a.c. mains voltage prior to application of the test level. τ  gbo Medizintechnik AG 2008 Version 1.3...
  • Page 38 80 MHz to 800 MHz 800 MHz to 2,5 GHz d=1,2√ √ √ √ P d=1,2√ √ √ √ P d=2,3√ √ √ √ P 0,01 0,12 0,12 0,23 0,38 0,38 0,73 Art.No. 019-7-0002  gbo Medizintechnik AG 2008 Version 1.3...