Guidance & Manufacturer's Declarations
Table 206
Portable and Mobile RF Communications
for ME Equipment and ME Systems that are
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Portable and Mobile RF Communications Equipment
The ClearChart 4P is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or user of the ClearChart 4P can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment and the ClearChart 4P as recommended below, according to
the maximum output power of the communications equipment.
Max Output
Power of Trans-
mitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d)
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects, and people.
13785-101 Rev. F
– Recommended Separation Distances between
Recommended Separation Distances for between
Separation (m)
150kHz to 80 MHz
d=(3.5/V1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
Equipment and the ClearChart 4P
NOT Life-supporting.
and the ClearChart 4P
Separation (m)
80 to 800 MHz
800MHz to 2.5GHz
d=(3.5/E1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
(continued)
Separation (m)
d=(7/E1)(Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.3333