  • Page 1 Thermo shaker PHMP/ PHMP-4 Operating instructions For version V.2GW...
  • Page 2: Table Of Contents

    Contents Safety..................3 General Information ..............5 Getting Started ................6 Operation of PHMP/PHMP-4 .............7 Specifications ................9 Guarantee and service............10 Version 2.03 - July 2013 Thermo shaker PHMP/PHMP-4 Page 2 Operating instructions...
  • Page 3: Safety

    Disconnect the external power supply unit from power socket to turn off the unit. If liquid penetrates into the unit, disconnect it from the external power supply unit and have it checked by a repair and maintenance technician. Thermo shaker PHMP/PHMP-4 Version 2.03 - July 2013 Operating instructions...
  • Page 4: Thermo Shaker Phmp/Phmp

    Do not check the temperature by touch. Use a thermometer. BIOLOGICAL SAFETY It is the user's responsibility to carry out appropriate decontamination if hazardous material is spilt on or penetrates into the equipment. Version 2.03 - July 2013 Thermo shaker PHMP/PHMP-4 Page 4 Operating instructions...
  • Page 5: General Information

    2. General Information The Thermo Shaker PHMP/PHMP-4 is designed for shaking a special platform that holds 1–4 standard microplates, and controlling the set temperature in the range of 25°C to 60°C (if the room temperature (RT) is less than 25°C).
  • Page 6: Getting Started

    (not less than 30 cm); Plug the external power supply unit into the socket at the rear side of the PHMP and position the unit so that there is easy access to the power switch and the external power supply unit.
  • Page 7: Operation Of Phmp/Phmp-4

    4. Operation of PHMP / PHMP 00:00 1000 35.0 STOP 0000 35.0 Actual hr : min Thermo-Shaker Fig. 1 Control panel Connect the external power supply unit to a grounded power socket and set the power switch located on the rear panel of the unit to position I ("ON").
  • Page 8 Place microplates on the platform and fix it with the special push-down clip by pressing it against the plate covers. PHMP-4: Unscrew the fixation screws. Place microplates on the platform and fix them by tightening the fixation screws. Caution! For model PHMP: Load only pairs of microplates for best fixing.
  • Page 9: Specifications

    5. Specifications 5.1 Thermo Shaker PHMP/PHMP-4 provides: soft or intensive sample shaking; · rotation speed control, stabilisation and display; · even shaking amplitude throughout the Thermo shaker platform; · required operation time setting and display; · automatic stopping of the platform movement after the set time expires;...
  • Page 10: Guarantee And Service

    When used in laboratory conditions and according to these working instructions, this product is guaranteed for TWO YEARS against faulty materials or workmanship. For full Details of the Grant Bio Warranty policy please contact Grant Instruments. Service For service, return for repair to our Service Department in the UK or, in other countries, to our distributor.
  • Page 11 Manufacturer: Equipment name/type number: Description of Equipment: Directive: Applied Standards Harmonized Standards: Thermo shaker PHMP/PHMP-4 Version 2.03 - July 2013 Page 11 Operating instructions...
  • Page 12 Grant Instruments (Cambridge) Ltd Shepreth Cambridgeshire SG8 6GB Tel: +44 (0) 1763 260811 Fax: +44 (0) 1763 262410 Email: scientificsales@grantinstruments.com www.grantinstruments.com Thermo shaker/PHMP/PHMP-4/18004/2.03...