X-Mind
DC
The "
" X-ray system is factory configured for "standard mode" operation which
determines:
®
X-Mind
DC
No. 2 "
" tube heads
An X-ray distance
SSD = 31cm
with the long 12" cone
An X-ray distance
SSD = 20cm
with the short 8" cone
with film type "D"
no. 1 CONTROL BUTTON
to make the exposure
work with
direct current at constant potential
work with
an X-ray voltage equal to 70kVp
work with
an X-ray current equal to 8mA
The above configuration depends on the position of 8 mini-switches (dip-switch
the timer electronic card:
DIP
SWICH
1
2
TYPE OF TUBEHEAD No. 1
3
TUBEHEAD No. 2
4
TYPE OF TUBEHEAD No. 2
5
CONTROL BUTTON No. 2
6
7
8
®
®
X
X
-
-
M
M
i
i
n
n
d
d
D
D
C
C

7. SYSTEM CONFIGURATION

** IF THE LONG 12" (31 cm) CONE IS USED **
** IF THE SHORT 8" (20 cm) CONE IS USED **
The installed tube head has the following features:
PARAMETER
TUBEHEAD No. 1
CONE
NOT AVAILABLE
NOT AVAILABLE
pressing the RX button
panel causes the LED 1 button to light up,
pressing the button again causes the LED 2 to
light up
on the control panel
on the control panel
on the control panel
The timer includes a key with an extension
cable
on the control panel must be lit
on the control panel must be lit
on the control panel must be lit
ON
INSTALLED
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
INSTALLED
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
INSTALLED
LONG (12")
NOT AVAILABLE
NOT AVAILABLE
on the control
LED 12" is lit
8" LED is lit
"D" LED is lit
the "DC" LED
the "70 kVp" LED
the "8 mA" LED
) on
OFF
NOT INSTALLED
ac
ALTERNATING CURRENT
SINGLE-PHASE
NOT INSTALLED
ac
ALTERNATING CURRENT
SINGLE-PHASE
NOT INSTALLED
SHORT (8")
3
5
3
5