SBS -2510 User Manual

Digital density meter
Hide thumbs Also See for SBS-2510:

Advertisement

Quick Links

SBS-2510
Digital Density Meter
User Manual
1-800-554-2243
www.sbsbattery.com
Test@sbsbattery.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SBS-2510 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SBS SBS-2510

  • Page 1 SBS-2510 Digital Density Meter User Manual 1-800-554-2243 www.sbsbattery.com Test@sbsbattery.com...
  • Page 3: Table Of Contents

    Perform the density determination ............13 Transfer Results (printing, exporting, pairing with SBS-6500) 6.1 Saving and exporting data................14 6.1.1 USB stick data transfer option..............14 6.1.2 Bluetooth export of results to SBS-6500 battery resistance tester……. Maintenance 7.1 Maintenance schedule..................7.2 Clean the SBS-2510 ..................
  • Page 4: Introduction

    SBS-2510 ..................... 21 Measurement ....................22 1 Introduction Thank you for choosing a SBS SBS-2510. The SBS-2510 is a handheld measuring instrument used to determine the density of liquids. About this document This document provides you with the information you need to get started with your SBS handheld SBS-2510.
  • Page 5: Definitions Of Signal Words And Warning Symbols

    2.2 Product specific safety notes Intended use The SBS-2510 is designed to be used by trained staff and is intended for measuring the density of liquid samples that are compatible with the materials with which they come into contact.
  • Page 6 CAUTION Injury due to incorrect handling of the battery The battery can explode or ignite and cause injury or damage the SBS-2510 if it is not handled correctly. 1 Do not store or use the battery at temperatures below -20 °C or above 60 °C.
  • Page 7: Design And Function Instrument Overview

    3 Design and Function Instrument overview No. Name Function 1 USB-C socket To connect the SBS-2510 with the AC adapter or a computer. 2 Battery cover To access the battery. 3 USB-A socket To connect a USB printer or Bluetooth dongle.
  • Page 8: User Interface

    Icons on the screen Icon Location Explanation Status ribbon Barcode reader is scanning. Status ribbon SBS-2510 is sending data to the printer. Status Shows the charge of the battery. ribbon Battery is charging. Status Syringe mode activated. The sample pump is deactivated.
  • Page 9: Installation And Commissioning

    Power Supply and Adapter 2510-PWRSUP USB-Charge Cable 2510-CABLE Bluetooth Dongle 2510-DONGLE User Manual – 4.2 Unpack the SBS-2510 1 Remove the SBS-2510 from the protective packaging. 2 Store the packing material for later transport over long distances. SBS-2510 Digital Density Meter...
  • Page 10: Charge The Sbs-2510

    Connect the SBS-2510 to the power supply 1 Insert the USB-C plug of the USB-C cable in the USB-C socket of the SBS-2510. 2 Install the cables in such a way that they cannot be damaged or interfere with operation.
  • Page 11: Setup For Working With The Sample Pump

    3 Insert the USB-A plug in the USB-A socket of the computer.  The SBS-2510 starts charging and the icon is displayed. Setup for working with the sample pump 4.4.1 Activate the sample pump mode 1 Press the Menu key.
  • Page 12: Operation 5.1 Start Up The Sbs-2510

     A short beep is played.  A welcome message is displayed.  The home screen opens. The SBS-2510 is ready to use. 5.2 Shut down the SBS-2510 − Press and hold any of the keys or the navigation joystick for more than 3 s.
  • Page 13: Typical Phases Of Density Determinations

    100 mPa*s. Syringes are typically used if the viscosity of the samples is above 100 mPa*s. You can find more information on working with the syringe in the Reference Manual. https://www.sbsbattery.com/sbs-2510-digital-battery-hydrometer.html SBS-2510 Digital Density Meter...
  • Page 14: Prepare The Measuring Cell For The Next Sample

    The SBS-2510 prompts users to enter the sample ID before the density is measured. Fill cycle The SBS-2510 prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement. SBS-2510 Digital Density Meter...
  • Page 15: Perform The Density Determination

    None The result is not printed. 5.5.2 Perform the density determination Start the method The SBS-2510 is set up to work with the sample pump.   The piston is in the lowermost position.  The home screen is open.
  • Page 16: Transfer Results (Printing, Exporting, Pairing With Sbs-6500)

    6.1.1 USB stick data transfer option 1. Insert a compatible USB drive into the USB socket of the SBS-2510 Digital Density Meter. (USB drive must be formatted in the FAT32 file format) 2. Press the “RESULTS” button on the main screen of the unit.
  • Page 17: Bluetooth Export Of Results To Sbs-6500 Battery Resistance Tester

    5. Press the “Start” button. You will see a message appear “Connecting to the USB Flash Drive” and receive an audible beep when the transfer of data is completed. Note: This method exports all data stored in the memory of the SBS-2510 into a single file that can be accessed via Microsoft Excel.
  • Page 18: Clean The Sbs-2510

    Because the measuring cell always contains residue from the last sample, there is a risk that the measuring cell can be damaged by the sample. To prevent such damages, SBS recommends that you clean the measuring cell at the end of each sample series.
  • Page 19: Check The Measurement Accuracy

    9 Clean the end of the filling tube with a clean tissue. 7.3 Check the measurement accuracy To check the measurement accuracy of the SBS-2510, you need to configure and perform a test. You can find more information about the configuration of a test in the Reference Manual.
  • Page 20: Perform The Test

    The result is not printed. 7.3.1.2 Perform the test Start the test The SBS-2510 is set up to work with the sample pump. The measuring cell has been cleaned. The piston is in the lowermost position. 1 Press the Menu key.
  • Page 21: Measures If A Test Fails

     Clean the measuring cell Page 16 7.4 Replace the battery The SBS-2510 works with a rechargeable 2400 mAh Li-Ion battery that you can order from Storage Battery Systems, LLC. 1 Press on the lower part (1) of the battery cover and push the battery cover up.
  • Page 22: Prepare The Sbs-2510 For Storage

    4 Reinstall the cover of the sample pump. 7.6 Prepare the SBS-2510 for storage The SBS-2510 can be stored at temperatures of -4…+122 °F (-20…+50°C). But the lifespan of the battery is reduced when it is stored at high temperatures.
  • Page 23 (31°C), linear fall to 50 % at 104°F (40°C) Altitude Up to 6560 Feet (2000 m) above sea level Pressure Atmospheric pressure In interior spaces Overvoltage category Pollution degree IP code IP5X Storage temperatures Characteristic Value Instrument -4…122°F (-20°C…+50°C) SBS-2510 Digital Density Meter...
  • Page 24 0...72 °Bé Accuracy ±0.1 °Bé % w/w Temperature range 50…122°F (10…50°C) Concentration range 0.0. 100.0 % w/w Accuracy ±1.0 % w/w % v/v Temperature range 50…122°F (10…50°C) Concentration range 0.0.... % v/v Accuracy ±1.0 % v/v SBS-2510 Digital Density Meter...
  • Page 25 Characteristic Value Brix (d) % w/w Range 0.0. 85.0 % w/w Accuracy ±0.3 % w/w Plato Range 0.0. 85.0 °P Accuracy ±0.3 °P SBS-2510 Digital Density Meter...