SBS-2500 HYDROMETER - QUICK START INSTRUCTIONS

Turn On Unit

 Press and hold the esc key until display appears
 Set time and date – (press ok and esc buttons simultaneously, enter password, scroll down to
"set time and date" press ok, esc )

Testing Your Batteries

 Place sample tube in sample to be tested
 Squeeze trigger button slowly to fill measuring cell
 Press the ok/meas. – holds result... wait for beeping to stop and then result is stored.
Note: if you need to resample, press escape, empty and retest.
 Press the ok/meas. – results are saved
 Press drain button down slowly to empty measuring cell

Transferring the Results

 Install SBS-2500 software onto your computer (software will include simple instructions)
 Open SBS-2500 software on computer
 Press and hold  arrow till display reads- memory out – execute? (ALL)
 Align IR window link on hydrometer w/IR computer link – at a maximum distance of
approximately 6 inches
 Press ok/meas. and results are transferred

Cleaning Instructions

 Place sample tube in water, draw water into measuring cell to clean
 Clean measuring cell several times w/water as needed
Before putting unit in case – clean and wipe down outside of hydrometer free from any
standing liquids w/soft cloth
For the full SBS-2500 Instruction Manual, go to:
www.sbsbattery.com/PDFs/SBS-2500-Hydrometer-Users-Manual.pdf
N56 W16665 RIDGEWOOD DR.  MENOMONEE FALLS, WI 53051  TEL. (262) 703-5800  FAX (262) 703-3073
Email:
techsupport@sbsbattery.com
Web Site:
www.sbsbattery.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SBS-2500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SBS SBS-2500

  • Page 1: Cleaning Instructions

     Press drain button down slowly to empty measuring cell Transferring the Results  Install SBS-2500 software onto your computer (software will include simple instructions)  Open SBS-2500 software on computer  Press and hold  arrow till display reads- memory out – execute? (ALL) ...
  • Page 2 6 months, or as needed • Do calibrate once per year Do Not’s SBS-2500 Spare Parts • Do not drop the SBS-2500 hydrometer Tube/Washer (P/N: SBS-2500TUBE/WASH) • Do not work in an explosion / hazardous environment • Do not press on measuring cell window Spare Parts Kit: O-ring, sample pump, sample tube/ • Do not clean the measuring cell with concentrated NaOH...

Table of Contents