SBS-2500 HYDROMETER - QUICK START INSTRUCTIONS
Turn On Unit
Press and hold the esc key until display appears
Set time and date – (press ok and esc buttons simultaneously, enter password, scroll down to
"set time and date" press ok, esc )
Testing Your Batteries
Place sample tube in sample to be tested
Squeeze trigger button slowly to fill measuring cell
Press the ok/meas. – holds result... wait for beeping to stop and then result is stored.
Note: if you need to resample, press escape, empty and retest.
Press the ok/meas. – results are saved
Press drain button down slowly to empty measuring cell
Transferring the Results
Install SBS-2500 software onto your computer (software will include simple instructions)
Open SBS-2500 software on computer
Press and hold arrow till display reads- memory out – execute? (ALL)
Align IR window link on hydrometer w/IR computer link – at a maximum distance of
approximately 6 inches
Press ok/meas. and results are transferred
Cleaning Instructions
Place sample tube in water, draw water into measuring cell to clean
Clean measuring cell several times w/water as needed
Before putting unit in case – clean and wipe down outside of hydrometer free from any
standing liquids w/soft cloth
For the full SBS-2500 Instruction Manual, go to:
www.sbsbattery.com/PDFs/SBS-2500-Hydrometer-Users-Manual.pdf