Performance Tests
External Mux/Demux
7 Press
on the HP 37718A/19A and set up the
RECEIVE
MAIN SETTINGS
8 Press
on the HP 37718A/19A and set up the
RECEIVE
display as shown below.
SETTINGS
9 Press
RUN/STOP
the Test Set and ensure that NO TROUBLE is displayed.
10 Press SINGLE error add key on the Test Set and ensure the Bit Error
count increments by one each time the key is pressed.
3-48
display as shown below.
on the Test Set. Press
STRUCTURED
TROUBLE SCAN
RESULTS
on