Performance Tests
STM-0/STS-1 Transmitter Output Waveshape
2 Connect the 52/155 Mb/s 75Ω DATA OUT port to the Oscilloscope
CHAN 1, terminate in 75Ω.
3
Select CHAN 1 on the Oscilloscope and set to 1MΩ input impedance.
Load the Telecom Mask
4 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask:
a/ SHIFT (Blue Key) APPLICATION.
b/
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select STS1_93 in the
highlighted MSK FILES window.
d/
LOAD SETUP 8 M1M2
Compare Pulse with Mask
5 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask:
a/ RECALL 8
b/ DEFINE MEAS
c/
softkey
COMPARE
d/
softkey
TEST ON
e/
softkey
M1M2
f/
AFTER FAIL CONTINUE
g/ RUN
6 The Oscilloscope will automatically display and compare an isolated
positive pulse with the mask limits. A PASS message should appear on
the Oscilloscope.
softkey.
softkey.
softkey
3-67