HP 37718A Installation And Verification Manual page 62

Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

N OTE
Performance Tests
PDH/DSn Transmitter Output
Load the Telecom Mask
5 On the Oscilloscope, use the following key sequence to select and load
the required Telecom Mask:
a/ SHIFT (Blue Key) APPLICATION.
b/
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select 8MG703 in the highlighted
MSK FILES window.
d/
LOAD SETUP 8 M1M2
Compare Pulse with Mask
6 On the Oscilloscope, use the following key sequence to compare the
pulse with the mask:
a/ RECALL 8
b/ DEFINE MEAS
c/
softkey
COMPARE
d/
softkey
TEST ON
e/
softkey
M1M2
f/
AFTER FAIL CONTINUE
g/ RUN
7 The Oscilloscope will automatically display and compare an isolated
positive pulse with the mask limits. A PASS message should appear on
the Oscilloscope.
It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS
on the Mask.
8 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 2.133V and 2.607 Volts.
Negative Pulse
9 On the Oscilloscope, use the following key sequence to access the
inverted mask.
a/ SHIFT (Blue Key) DISK
b/
softkey
DIRECTORY
c/ Use ENTRY/MEASURE control to highlight INV_MASK
d/
CHANGE DIRECTORY
3-22
softkey.
softkey.
softkey
softkey

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniber 719Omniber 71837719a

Table of Contents