Woodpecker DTE Implant-X Instruction Manual

Woodpecker DTE Implant-X Instruction Manual

Dental implant device

Advertisement

Guilin Woodpecker Medical Instrument Co., Ltd.
The pictures are only for reference. The final interpretation rights belong to Guilin Woodpecker Medical Instrument Co., Ltd. Woodpecker reserves the right to change
the design of the equipment, the technique, fittings, instruction manual and the content of the original packing list at any time without further notice.
The product were authorized appearance patent, counterfeit will be sued.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DTE Implant-X and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Woodpecker DTE Implant-X

  • Page 1 Guilin Woodpecker Medical Instrument Co., Ltd. The pictures are only for reference. The final interpretation rights belong to Guilin Woodpecker Medical Instrument Co., Ltd. Woodpecker reserves the right to change the design of the equipment, the technique, fittings, instruction manual and the content of the original packing list at any time without further notice.
  • Page 2 Thank you for purchasing Woodpecker Dental Implant Device Implant-X. In order to guarantee the correct operation, it is recommended to read this instruction manual carefully before operation. For convenient reading, it is recommended to put it where is available at any time.
  • Page 3: Table Of Contents

    Contents 1. Product introduction ......................1 2. Accessories description ....................3 3. Control of host and foot pedal ..................4 4. Installation ........................7 5. Operation ........................10 6. Clean, disinfection, and sterilization ................19 7. Error code and solution (error alarm interface) ............21 8. Storage and maintenance .....................21 9.
  • Page 4: Product Introduction

    1. Product introduction 1.1 Precautions Danger 1. To prevent electric shock, do not use wet hands to pull the power cord; be sure to prevent the control circuit from water; use a grounded electrical outlet. 2. Keep it away from explosives and combustibles, with special care not to use this machine for patients who use nitrous oxide anesthesia.
  • Page 5 Woodpecker reserves the right to change the design of the equipment, the technique, fittings, instruction manual and the content of the original packing list at any time without further notice. The pictures are only for reference. The final interpretation rights belong to Guilin Woodpecker Medical Instrument Co., Ltd.
  • Page 6: Accessories Description

    1.4 Safety requirement Guilin Woodpecker Medical Instrument Co., Ltd. will NOT be responsible for any direct or indirect damages and losses under the following conditions: The equipment is used for any purpose that is not mentioned in the scope of use.
  • Page 7: Control Of Host And Foot Pedal

    3. Control of host and foot pedal 3.1 Control of host keys 3.1.1 Working interface and control of keys...
  • Page 8 1. Light key Motor LED switch; circular touch to control the on/off of LED. 2. Store key Store key; click to store setting specifications. 3. Water control key Click to select water volume. Six water gear, including 00%, 20%, 40%, 60%, 80%, and 100% 4.
  • Page 9 3.1.3 Error alarm interface Error 01 The foot is not connected! Please check whether the foot is in good contact. Figure 3 Error alarm interface As shown in Figure 3, the warning 0x indicates alarm number. Please refer to section 7 Error code and solution for specific number and the corresponding content.
  • Page 10: Installation

    level 6, pressing once to loops back to level 1. 2. Program switchover button Used for choosing needed programs. The program is changed each time the button is stepped. At the Program 8, pressing once to loops back to Program 1. 3.
  • Page 11 4.2 Accessories connection Figure 4 4.2.1 Installation of foot pedal: Connect the foot pedal plug to the foot pedal socket and tighten those two fixing screws (Figure 4 - E) 4.2.2 Power cord installation: Plug the power cord output into the power supply socket of the device (Figure 4 - D) 4.2.3 Installation of infusion bottle holder: Insert the infusion bottle holder into the fixing hole on the right rear of the shell;...
  • Page 12 4.2.6 Peristaltic pump tube installation: A.Turn the peristaltic pump knob clockwise to the "OPEN" indicator and open the pump head (Figure 5-A). B.Place the tube in the impeller of the peristaltic pump (Figure 5 - B). C.Turn the peristaltic pump knob clockwise to the "CLOSE" position and close the pump head (Figure 5 - D). 4.2.7 Complete machine effect after installations of all accessories: (Figure 6) Figure 6 4.2.8 Switch on (Figure 4 - C) ;...
  • Page 13: Operation

    5. Operation 5.1 Program 5.1.1 Choice of programs Implant-X owns 8 programs. There are two ways to choose program: 1. Touch the corresponding icons on the screen. 2. Step on the “Program switchover” button on the foot pedal. 5.1.2 Function description of programs The function of each program is as shown as follow: Icon Function...
  • Page 14 Hole-drilling Determine the direction and depth of hole-drilling. Hole- Determine the diameter of the hole. broadening Tapping Make a thread on the hole to match the implant. Implanting Implant dental implants into alveolar bone. Lock the Screw the nut onto the dental implant. abutment screw User defined Change straight handpiece, contra-angle with different speed ratios for different...
  • Page 15 The range of different parameters and their factory settings are as shown in the table below: Icon Function Speed/ rpm Torque/ N·cm Speed ratio Water output/% Positioning 200-2500 1000(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) Hole-drilling 200-2500 800(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) Hole-broadening 200-2500 600(D) 5-80 35(D)
  • Page 16 5.2 Default parameter adjustment Within specified range, the adjustable parameters are as follow: 1. Maximum speed 2. Maximum torque 3. Water output 4. Speed ratio 5.2.1 Maximum speed adjustment Touch the “Speed” (+, -) key to adjust motor speed. The speed will change each time after touching the “Speed” key. Long press the “Speed”...
  • Page 17 5.2.2 Maximum torque adjustment Touch the “Torque” (+, -) key to adjust maximum torque output of motor. The torque will change each time after touching the “Torque” key. Long press the “Torque” key to accelerate the change of torque setting value. 5.2.3 Water output adjustment...
  • Page 18 Touch the “Water volume” key on the screen to adjust. There are 6 water levels. The water level will change to the next level after each touch. Step on the blue “Water volume adjustment” button to adjust water volume. 5.2.4 Speed ratio adjustment Adjust by pressing the "Speed Ratio"...
  • Page 19 5.3 Motor rotating direction adjustment Touch the key shown above to change the rotating direction of motor. Step on the “Forward/reverse rotation” during operation to change the rotating direction of motor.
  • Page 20 5.4 Motor LED adjustment Touch the “LED” button to for setting to determine the on or off state of LED while stepping on foot pedal. The state of LED will change once after each touch. Only the device with LED owns this function. 5.5 Save the parameters...
  • Page 21 After finishing the above steps, press “Store” key. You will hear a beep if the parameters are saved. 5.6 Standard operation 1. After installation of corresponding accessories, connect to the power supply, and turn on the power supply. After booting, the displayed interface is default to be Program 1.
  • Page 22: Clean, Disinfection, And Sterilization

    5. Torque protection will start as the torque reaches preset value. Meanwhile, the motor slows down to stop, preventing from generating excessive torque. Release foot pedal to remove torque protection. Step again, and the motor will rotate under preset torque value.
  • Page 23 Warning: Clean and disinfect the main unit and foot controller regularly. When subjecting the main unit and foot controller to cleaning and disinfection ensure that the charging cable is not connected and that the charging socket is closed. Warning: Only the following parts can be sterilized: Handpiece, motor and its tail cable, holder of silicone handpiece, motor disinfection stopper, motor cable-cooling pipe clamp, and lubricant nozzle.
  • Page 24: Error Code And Solution (Error Alarm Interface)

    Seal the motor, motor tail cord and motor holder Sterilization bag should be in the sterilization bag. in accordance to applicable Packaging —— sterilization standard, and is suitable for sterilization. Put the sterilization bags into the The sterilized items autoclave sterilizer. It is recommended should cool down to room Sterilization ——...
  • Page 25 8.2 Do not store the machine together with articles that is poisonous, combustible, caustic, or explosive. 8.3 This device should be stored in a room where the relative humidity is not more than 10%~93%, atmospheric pressure is 70kPa~106kPa, and the temperature is-20℃ ~ +55℃. 8.4 Turned off power switch and unplug the power plug when the device is not in use.
  • Page 26: Symbols

    9. Symbols Mark Follow Instructions for Use Use indoor only Alternating current Socket for the foot switch Caution mechanical injury Manufacturer Date of manufacture Drip-proof Strong water spraying experiment Type B applied part Can be autoclaved Caution Protective earthing Serial number Humidity limit for storage Atmospheric pressure for storage Temperature limit for storage...
  • Page 27: Specifications

    10. Specifications 10.1 Host specifications Model: Implant-X Device for intermittent operation: 3 min ON,10 min OFF Power supply voltage: 200-240V~ Power supply frequency: 50/60Hz Software version: Implant-V1.0.0 Input power: 150VA Fuses: 2×T1.6AL 250V Applied parts: Contra-angle handpiece Maximum Temperature: 41.8℃ Maximum water volume: 110ml/min Dimension: 265mm*263mm*119mm Operation environment:...
  • Page 28: After-Sales Service

    13. Statement Woodpecker reserves the right to change the design of the equipment, the technique, fittings, instruction manual and the content of the original packing list at any time without further notice. The pictures are only for reference. The final interpretation rights belong to Guilin Woodpecker Medical Instrument Co., Ltd.
  • Page 29: Emc-Declaration Of Comformity

    14.6 Warnings To request a warranty, please send your device, warranty card, and invoice for the device to your Woodpecker distributor / importer within the warranty period. In order to be repaired during the warranty period, the purchaser shall return the repaired product to the distributor / importer at their expense.
  • Page 30 Emissions test Compliance Electromagnetic environment - guidance RF emissions Group 1 The model Implant-X uses RF energy only for its internal function. Therefore, its RF CISPR 11 emissions are very low and are not likely to cause any interference in nearby electronic equipment.
  • Page 31 Voltage dips, short <5 % UT <5 % UT Mains power quality should be that of a typical interruptions and voltage (>95% dip in UT.) (>95% dip in UT.) commercial or hospital environment. If the user of the variations on power supply for 0.5 cycle for 0.5 cycle models Implant-X requires continued operation during...
  • Page 32 Conducted RF 3 Vrms Portable and mobile RF communications equipment should be used no closer to any lEC 61000-4-6 150 kHz to 80 MHz part of the models Implant-X, including cables, than the recommended separation Conducted RF 6 Vrms 3V/m distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 33 The model Implant-X is intended for use in electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the model Implant-X can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model Implant-X as recommended below, according to the maximum output power of the communications equipment.
  • Page 34 ZMN/WI-09-426 V1.3-20181218...

Table of Contents