Setup Of Capacity (Maximum Load) - KERN HFM Service Manual

Crane scale
Table of Contents

Advertisement

3.3 Setup of capacity (maximum load)

1)
Switch off the balance.
2)
Open the top cover of housing.
3)
Set jumper on pin J2.
4)
Switch on the balance and during the self test press the ZERO and the
button once at the same time.
P3
- - - " is displayed.
5)
Enter password „
Pressing the
digit by one.
Pressing the HOLD button moves the digit.
6)
The current capacity is displayed, e.g. 1000
Press
button.
7)
Now balance will return to normal weighing mode.
8)
Switch off the balance and remove the jumper J2.
8
999
" and confirm with ZERO button.
button increases the numerical value of the blinking
button to change the capacity and confirm with ZERO
HFM-SH-e-1110

Advertisement

Table of Contents
loading