Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 2
SCC-CO20 Isolated Current Output Module The SCC-CO20 is an isolated current output module that outputs up to 20 mA. The output current level is proportional to the voltage applied to the module by an analog output channel of an E/M Series DAQ device.
SC-2345 refers to both the SC-2345 connector block and the SC-2345 with configurable connectors. SCC refers to any SCC Series signal conditioning module. What You Need to Get Started To set up and use the SCC-CO20, you need the following items: ❑ Hardware –...
Plug the SCC-CO20 into either SC-2345/2350 analog output socket J(X+17), where X is 0 or 1. The SC-2345/2350 routes the voltage output of E/M Series DAQ device analog output channel AO (X) to the SCC-CO20. Connecting the SCC-CO20 to a Load Note The signal names have changed.
Page 5
Figure 1. SCC-CO20 Signal Connections to E/M Series DAQ Device Using the Isolated Loop Supply of the SCC-CO20 You can connect the SCC-CO20 to a load using either the isolated internal loop supply or an external supply. The internal loop supply outputs a voltage of 15 V with respect to the isolated ground reference.
Page 6
If your load is grounded and you are using the internal supply, connect the load ground to the current sink terminal at pin 2 on the SCC-CO20. Do not connect the load ground to the isolated ground terminal because doing so could defeat the module isolation.
Page 7
ISO GND SCC-CO20 SUPPLY 15 V iso Load – SINK Figure 3. Connecting Two SCC-CO20 Modules to a Higher-Resistance Load SCC-CO20 Isolated Current Output Module User Guide ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 8
– b. Referenced load and external voltage supply (two circuits) Figure 4. Connecting the SCC-CO20 Using an External Loop Supply When you use an external loop supply, the order of the supply and load does not matter as long as you do not create a second loop through which current flows.
Page 9
Calibrating Gain and Offset Errors The SCC-CO20 is calibrated at the factory before shipment. If you want to adjust the gain and offset of the SCC-CO20 in your system using an E/M Series DAQ device, you need a current meter (ammeter) with an accuracy of at least ±2 μA.
Page 10
Normally you would adjust to 0 μA in order to calibrate the offset, but because Note the SCC-CO20 cannot source current, it is impossible to know when you have adjusted past 0 μA. Using the SCC-CO20 The SCC-CO20 is a 0 to 20 mA programmable current sink. The current output is controlled by AO (X), a 0 to 10 V unipolar DAC output channel of an E/M Series DAQ device.
+15 V ..........5.8 mA max –15 V ..........5.8 mA max Digital power (+5 V) ......645 mW +5 V ..........129 mA max SCC-CO20 Isolated Current Output Module User Guide ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
Page 15
E/M Series AO GND — E/M Series AO GND — — — — +5 V — +5 V REF +15 V SCC-CO20 Isolated Current Output Module User Guide ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 16
— National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Page 17
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...