Fluke ScopeMeter 123 Service Manual page 69

Industrial scopemeter
Hide thumbs Also See for ScopeMeter 123:
Table of Contents

Advertisement

15. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is
triggered. To repeat the test, start at step 12.
16. Set the 5500A to Standby.
17. Press
to clear the display.
18. Select the following test tool setup:
Press
to open the SCOPE INPUTS menu.
Press
to open the TRIGGER menu, and choose:
I
INPUT:
B | SCREEN UPDATE:
Press
to enable the arrow keys for Trigger Level and Slope adjustment.
Using
select positive slope triggering (trigger icon
Using
positive slope triggering, the trigger level is the top of the trigger icon ( ).
19. Set the 5500A to source 0.4V DC.
20. Verify that no trace is shown on the test tool display, and that the status line at the
display bottom shows
then press
to re-arm the test tool for a trigger.
21. Increase the 5500A voltage slowly in 0.1V steps, using the 5500A EDIT FIELD
function, until the test tool is triggered, and the traces are shown.
22. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is
triggered.
To repeat the test, start at step 19.
23. Set the 5500A to Standby.
24. Press
to clear the display.
25. Press
to enable the arrow keys for Trigger Level and Slope adjustment.
26. Using
select negative slope triggering ( ).
27. Using
set the trigger level to +2 divisions from the screen center. For
negative slope triggering, the trigger level is the bottom of the trigger icon ( ).
28. Set the 5500A to source +3V DC.
29. Verify that no trace is shown on the test tool display, and that the status line at the
display bottom shows
then press
to re-arm the test tool for a trigger.
30. Decrease the 5500A voltage in 0.1V steps, using the 5500A EDIT FIELD function,
until the test tool is triggered, and the traces are shown.
31. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is
triggered. To repeat the test, start at step 28.
32. When you are finished, set the 5500A to Standby.
I
FREE RUN | AUTO RANGE:
set the trigger level to +2 divisions from the screen center. For
. If the display shows the traces and status
Wait:B
. If the display shows the traces and status
Wait:B
Performance Verification
4.5 Input A and Input B Tests
I
>15HZ
).
,
Hold:B
,
Hold:B
4
4-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents