7. Start-up

The cartridges must be at room temperature when they are inserted.
The cartridge spike is sharp.
Be careful not to injure yourself on the projecting spike.
Whenever you replace a cartridge:
>
always clean and disinfect the cartridge compartment,
>
perform the initial filling.
Observe the manufacturer's instructions on use of the disinfectant.
18
 Remove the cartridge
compartment cover.
Inserting a cartridge.
> Push the cartridge in as far
as it will go.

Insertion/removal of a cartridge

e Removing a cartridge.
> Pull out the cartridge.
Replace the cartridge
compartment cover.