Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual.
Page 5
Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Page 6
RIGOL battery) may cause explosion. Use RIGOL specified battery only. Handling Safety. Please handle with care during transportation to avoid damage to buttons, knob interfaces and other parts on the panels. DG4000 Performance Verification Guide...
RIGOL Safety Terms and Symbols Terms Used in this Manual. These terms may appear in this manual: WARNING Warning statements indicate conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate conditions or practices that could result in damage to this product or other property.
RIGOL Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben. Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel.
Page 9
Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch RIGOL autorisiertem Personal durchgeführt wird. Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen am Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal durchgeführt werden. Belüftung sicherstellen.
Page 10
Wenneine Batterieverwendet wird, vermeiden Sie hohe Temperaturen bzw. Feuer ausgesetzt werden. Bewahren Sie es außerhalbder Reichweitevon Kindern auf. UnsachgemäßeÄnderung derBatterie (Anmerkung: Lithium-Batterie) kann zu einer Explosion führen. VerwendenSie nur von RIGOL angegebenenAkkus. Sicherer Transport. Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.
RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUTION Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die Schäden am Gerät hervorrufen können.
RIGOL Document Overview This manual guides users to correctly test the performance of RIGOL DG4000 series function/arbitrary waveform generator. For the operation methods of the instrument, please refer to the DG4000 User’s Guide. Main topics in this manual: Chapter 1 Overview This chapter introduces the preparations and precautions of the performance verification test.
Contents RIGOL Contents Guaranty and Declaration ................. I General Safety Summary ................ II Safety Terms and Symbols ............... V Allgemeine Sicherheits Informationen ........... VI Sicherheits Begriffe und Symbole ............IX Document Overview ................. X Chapter 1 Overview ................1-1 Test Preparations ................... 1-1 Recommended Test Devices ..............
Table 1-1 Recommended Test Devices Device Performance Requirement Recommended Frequency >10 MHz Agilent 53131A Counter Accuracy: 0.1 ppm RIGOL DM3068 Digital Multimeter 6 -30 dBm to +20 dBm Power Meter Accuracy: ±0.02 dB Agilent E4416A Resolution: 0.01 dB Spectrum RIGOL DSA1030A Minimum resolution bandwidth is 10 Hz.
The specifications of each test item are provided in chapter 2. For other specifications, refer to DG4000 User’s Guide or DG4000 Data Sheet (can be downloaded from www.rigol.com). Tip: All the specifications are only valid when the instrument has been warmed up for more than 30 minutes under the specified operation temperature (18℃...
Chapter 2 Performance Verification Test RIGOL Chapter 2 Performance Verification Test Taking DG4162 as an example, this chapter introduces the method of performance verification test for DG4000 series function/arbitrary waveform generator. Test items include: Frequency Accuracy Test AC Amplitude Accuracy Test ...
CH1 as an example; the test method is also applicable to CH2) of DG4000 with the signal input terminal of the frequency counter using the dual-BNC cable as shown in the figure below. DG4000 Series Frequency Counter ◎ Figure 2-1 Connect DG4000 and the Frequency Counter Press Preset at the front panel of DG4000 to restore the instrument to its default setting.
Page 19
Chapter 2 Performance Verification Test RIGOL frequency counter and check if the readings are between 0.999998 MHz and 1.000002 MHz. Repeat steps 1 through 4 to verify the frequency accuracy of CH2. DG4000 Performance Verification Guide...
CH2) of DG4000; connect the 50Ω load and the voltage input terminals of the digital multimeter using the BNC - dual banana cable as shown in the figure below. DG4000 Series ◎ ◎ ◎ ◎ 50 Ω Load Figure 2-2 Connect DG4000 and the Digital Multimeter via the 50Ω...
Page 21
Chapter 2 Performance Verification Test RIGOL Record the reading of the DMM and check if the reading exceeds the range listed in the Limit of Actual Output Amplitude (Vrms) in Table 2-1. Keep other settings of DG4000 unchanged, set the amplitude to 100 mVpp, 500...
RIGOL Chapter 2 Performance Verification Test DC Offset Accuracy Test Specification: Output Characteristics Offset (into 50 Ω) Accuracy ±(1% of setting + 5mV + 0.5% of amplitude) Test Procedures: Make sure that DG4000 has been warmed up for 30 minutes. Connect the 50Ω...
Page 23
Chapter 2 Performance Verification Test RIGOL Table 2-2 Limits of DC Offset Accuracy Test Setting Setting Offset Permitted Error Limit of Actual Offset Amplitude -2.5 V ±0.005 V -2.505 V ~-2.495 V -1 V ±0.02 V -1.02 V ~-0.98 V -500 mV ±0.025 V...
RIGOL Chapter 2 Performance Verification Test AC Flatness Test Specification: Output Characteristics Typical Flatness ≤10 MHz: ±0.1 dB (relative to 1kHz Sine wave, ≤60 MHz: ±0.2 dB 500 mVpp, 50 Ω) ≤100 MHz: ±0.4 dB ≤160 MHz: ±0.8 dB Test Procedures: Make sure that DG4000 has been warmed up for 30 minutes.
Page 25
RIGOL c) Disable power reference. Disconnect the signal generator and the multimeter. Connect the power sensor with the channel output terminal of DG4000 via the BNC (f)-N (m) connector. DG4000 Series Power Meter ◎ Power Sensor Figure 2-3 Connect DG4000 and the Power Meter Keep other settings of DG4000 unchanged, set the frequency to 10 MHz.
N-BNC adaptor as shown in the figure below. Press Preset at the front panel of DG4000 to restore the instrument to its default setting. DG4000 Series Spectrum Analyzer ◎ Figure 2-4 Connect DG4000 and Spectrum Analyzer Configure DG4000: the output impedance of the channel tested is 50 Ω...
Page 27
Chapter 2 Performance Verification Test RIGOL Enable the peak table function of the spectrum analyzer and set appropriate peak measurement parameters to make the peak table display the measurement values of the fundamental wave and the second harmonic. Record the measurement values in the peak table and check if the harmonic distortion is...
Page 28
RIGOL Chapter 2 Performance Verification Test Note : The Second Harmonic Distortion = The Measurement Value of the Second Harmonic - The Measurement Value of the Fundamental Wave The fundamental wave is also considered as the first harmonic. For example, if the measurement value of the fundamental wave is -10 dBm and the measurement value of the second harmonic is -72 dBm, the second harmonic distortion = (-72) –...
CH2) of DG4000 with the signal input terminal of oscilloscope using the dual-BNC cable as shown in the figure below. Press Preset at the front panel of DG4000 to restore the instrument to its default setting. DG4000 Series Oscilloscope ◎...
Page 30
RIGOL Chapter 2 Performance Verification Test Set the edge type to falling edge. Record the measurement result of the fall time and check if it is within the specification range. Repeat steps 1 through 5 to verify the rise/fall time of CH2.
Chapter 2 Performance Verification Test RIGOL Spurious Signal Test Specification: Sine Wave Spectrum Purity Spurious signal Typical (0 dBm) (non-harmonic) ≤10 MHz: <-65 dBc >10 MHz: -65 dBc+6 dBc/octave Test Procedures: Connect CH1 of DG4000 with the RF input terminal of the spectrum analyzer using the dual-BNC cable and N-BNC connector as shown in Figure 2-4.
Page 32
RIGOL Chapter 2 Performance Verification Test Keep other settings of the spectrum analyzer unchanged, set the stop frequency of the spectrum analyzer to 150 MHz, 300 MHz, 500 MHz and 600 MHz respectively (correspond to the output frequency of DG4000) and repeat step 5.
Chapter 2 Performance Verification Test RIGOL Overshoot Test Specification: Signal Characteristics Square Typical (100 kHz, 1 Vpp) Overshoot <3% Test Procedures: Connect CH1 of DG4000 and the signal input terminal of the oscilloscope using dual-BNC cable as shown in Figure 2-5. Press Preset at the front panel of DG4000 to restore the instrument to its default setting.
Appendix RIGOL Appendix Test Record Form RIGOL DG4000 Series Function/Arbitrary Waveform Generator Performance Verification Test Record Form Model: Tested by: Test Date: Frequency Accuracy Test Channel: CH1 Measurement Limits calculated from Waveform Settings Pass/Fail Values specification Sine Frequency: 1 0.999998 MHz...
Page 37
Appendix RIGOL DC Offset Accuracy Test Channel: CH1 Setting Measurement Limits calculated from Settings Pass/Fail Offset Values specification -2.5 V -2.505 V ~-2.495 V Frequency: -1 V -1.02 V ~-0.98 V 1 kHz -500 mV -0.525 V ~-0.475 V Amplitude: -0.030 V...
Page 38
RIGOL Appendix AC Flatness Test Channel: CH1 Setting Measurement Settings specifications Pass/Fail Frequency Values 1 kHz Amplitude: 10 MHz ±0.1 dB 500 mVpp 60 MHz Offset: 0 V ±0.2 dB Impedance: 100 MHz ±0.4 dB 50 Ω 160 MHz ±0.8 dB...
Need help?
Do you have a question about the DG4000 Series and is the answer not in the manual?
Questions and answers