Removing The Holster - Fluke 150 Series Calibration Manual

Insulation testers
Hide thumbs Also See for 150 Series:
Table of Contents

Advertisement

150X Series
Calibration Manual
Holster
Battery Door
Case
Screws
Case
Bottom

Removing the Holster

The standard Tester comes equipped with a snug-fitting yellow rubber holster. The
holster helps protect the Tester from rough handling and normally remains on the Tester.
The first step in disassembling the Tester is to remove the holster.
Use the following procedure to remove the holster:
1. Looking at the Tester, place your thumbs on the top corners of the holster and firmly
2. Using both thumbs, push the holster up and over both corners of the Tester.
3. Continue pushing on the holster until both of its inside corners are resting on top of
4. Rest the heel of one hand behind the display, and place all four fingers of the same
5. Firmly grasp the Tester with the other hand, and using your fingers, peel the holster
6. Slide the Tester up and out of the holster.
To install the holster, position the bottom of the Tester in the bottom of the
holster, and press the top of the instrument firmly into of the holster.
10
Figure 3. Disassembling the Tester
grasp the Tester.
the Tester.
hand along the upper front edge of the holster.
over the top of the Tester.
riešenia na presné meranie
Fuse
Plastic Shields
Red
Black
Note
http://www.elso.sk
PCA Screws
PCA
Light-dispersing
Back Panel
Glass LCD
Display
Elastomeric
Strip
Gray Plastic
Bezel
Keypad
Case Top
ecz20f.eps