7.2
KERN HUS
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Switch-off balance and attach a carrying help if necessary.
Press and hold ON/ZERO- and MODE whilst at the same time switching on the
scales with the carrying aid attached. Release keys, „CAL?" will be displayed.
Press the MODE-key, „YES" will be displayed, followed by an alternating display
such as "Load 100 or 40", see table 1.
The admissible adjustment weights will be displayed.
Attach the admissible adjustment weights (s. table 1). Wait until the scales display
„done", and start to count down.
After successful adjustment the balance carries out a selftest and automatically
returns to weighing mode. Remove adjustment weight and also the carrying aid, if
attached.
An adjusting error or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.
HUS/HUM-BA-e-1131
39