6.2

Performing a factory reset (reset to factory defaults)

A factory reset can also be performed via the Web Interface (see section "Performing a
factory reset" of the TICRO 100 User Manual).
Performing a factory reset may possibly result in an IP address change of TICRO 100 due to
automatic IP address assignment by a DHCP server.
The reset pushbutton can be accessed through the hole in the rear panel using a thin
pointed object like the wire of a resistor with a diameter < 0.9 mm, for example.
Resetting the device to the factory defaults may be necessary if you forgot your password or if you
cannot access the TICRO 100 Web Interface anymore because you selected the wrong Ethernet
network port by mistake before.
1. Press the reset pushbutton and keep it pressed.
2. The green LED S1 goes off.
3. After approx. 5 s, the red LED S2 lights up together with the green LED S1. Keep the button
pressed and wait until the green LED S1 goes off (the red LED S2 is still on).
4. Release the reset pushbutton.
5. After approx. 15 s, the red LED S2 goes off and the green LED S1 lights up.
6. The device rebooted successfully and has now the factory default configuration settings.
OMICRON
Troubleshooting
23