TICRO 100 Quick Start Guide
Connector
12
Description
DC power input socket (MC 1.5/2-STF-3.81, connector supplied on delivery).
The correct polarity is printed on the front plate.
You can use this connector to connect a 18 ... 57 VDC supply voltage, either
as an alternative to the plug-in power supply unit supplied on delivery (DC
power input socket on rear side), or in addition to the plug-in power supply
unit to provide power supply redundancy for TICRO 100. For more details,
please refer to chapter
WARNING
Death or severe injury caused by high voltages possible.
► Use a power supply unit that complies with the SELV
standard if product safety according to IEC 61010-1 and
IEC 60255-27 is required. The plug-in power supply unit
supplied on delivery complies with the SELV standard.
NOTICE
Equipment damage possible.
The DC power inputs of TICRO 100 are protected against polarity reversal
if only one input is used at a time.
► When using both DC power inputs, make sure to observe the correct
polarity for both inputs. Reversed polarity on one of the DC power
inputs would make the reverse polarity protection ineffective and could
damage TICRO 100.
Power Supply
on page 16.