Heater Test; Uv Sterilization - RBC Bioscience MagCore Plus II Operation Manual

Hide thumbs Also See for MagCore Plus II:
Table of Contents

Advertisement

━ Heater Test

Please Select Heater Test Button.
Please put the thermometer into the Heat Block
Well 1 and press Next.
Select Test Temperature
Now the heater will heat up to 65° C or 90° C,
please check whether the temperature shown on
the thermometer is consistent with what's shown
on the screen.
The acceptable temperature range is 65± 5° C or
90± 5° C.
www.rbcbioscience.com
21

━ UV Sterilization

Please Select the Sterilization Button.
Please select the sterilization time and press Start
button.
Warning!
The UV lamp is dangerous. Please do not
look straight at the inner of the machine
when operating the machine.
Please stop operation before material
change.
Please do not open this cover by
untrained operators.
Please attention for UV light. Open.

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for RBC Bioscience MagCore Plus II

Table of Contents