RBC Bioscience MagCore Plus II Operation Manual page 20

Hide thumbs Also See for MagCore Plus II:
Table of Contents

Advertisement

C. Test the Tip of the W3
1. Please put the Pipette Tip into Well 3 of T-Rack.
2. Please add 1 ml water to the Sample Tube and
put it in the Well1 of the T-Rack.
3. Please select the Test W3 Tip.
4. Please check whether the liquid levels in all tips
are consistent with each other. If the liquid levels
are not consistent, please check the O-Ring
(Please refer to the Replace O-Ring chapter for
more information).
www.rbcbioscience.com
19
D. Test the Tip of the W4
1. Please put the 200μl SP Tip into Well 4 of T-Rack.
2. 2. Please add 1 ml water to the Sample Tube
and put it in Well1 of the T-Rack.
3. Please select the Test W4 Tip.
4. Please check whether the liquid levels in all tips
are consistent with each other. If the liquid levels
are not consistent, please check the O-Ring
(Please refer to the Replace O-Ring chapter for
more information).