Clean Piercing Needles; Eject Tip - RBC Bioscience MagCore Plus II Operation Manual

Hide thumbs Also See for MagCore Plus II:
Table of Contents

Advertisement

━ Clean Piercing Needles

Please select Piercing Needles Button.
Take Cartridge Rack Out
1. Please prepare a soft cloth sprayed with 70%
alcohol.
2. Please clean the Piercing Needles with the cloth.
Warning!
The tips of the Piercing Needles
are extremely sharp. Please be
careful when you clean them.

━ Eject Tip

If the system is stopped from power shortage or
other abnormal conditions, and the tips are still on
the pistons, you have to press Reject Tip to remove
the tips from the pistons.
Please select Reject Tip Button.
or
MagCore® Plus II
20