DESCRIPTION
The new FLS M9.07 is a dual monitor which combines conductivity and flow
measurements. A 4" wide full graphic display shows measured values clearly
together with many other useful information. Moreover, due to a multicolor
display plus a powerful backlight, measurement status can be determined
easily from afar also. A tutorial software guarantees a mistake-proof and fast
set up of every parameters. Different type of calibrations can be performed
to fit user needs for both measurements. A 4-20mA output dedicated to
each measurement grants to remote values to a external device. A proper
combination of digital outputs allows customized setups for any process to be
controlled.
CONNECTIONS TO INSTRUMENTS
F3.00
M9.07
X
ULF
M9.07
X

TECHNICAL DATA

General
• Associated sensors: FLS conductivity sensors & FLS hall effect flow sensors
or FLS F6.60 flow magmeters
• Materials:
- case: ABS
- display window: PC
- panel & wall gasket: silicone rubber
- keypad: 5-button silicone rubber
• Display:
- LC full graphic disply
- backlight version: 3-colours
- backlight activation: User adjustable with 5 levels of timing
- update rate: 1 second
- enclosure: IP65 front
• Conductivity input range: 0,055÷200000 μs/cm (Cell Constant: 0.01 - 10.0)
• Conductivity measurement accuracy: ± 2.0 % of reading value
• Temperature input range: -50÷150°C (-58÷302°F) (with Pt100-Pt1000)
• Temperature measurement resolution: 0,1°C/°F (Pt1000); 0,5°C/°F (Pt100)
• Flow input range (frequency): 0÷1500Hz
• Flow input accuracy (frequency): 0,5%