MAX MODE (icon reports MAX)
Cond
Output relaxed
Output energized
FLOW
MIN MODE (icon reports MIN)
Flow
Output relaxed
Output energized
WINDOW MODE (icon reports WDW)
Flow
Output relaxed
Output energized

SOFTWARE UPDATING

In order to update the Instrument Software with a New Firmware Release
follow the suggested procedures:
TO UPDATE INSTALLED UNITS
- Download the interface software "FLS Calibration System" and the Updated
Software on www.flsnet.it
- Launch the software "FLS Calibration System" on the laptop
- Select OPTION and then UPGRADE FIRMWARE
- Confirm the "Firmware Upgrade" procedure by ENTER
- Connect M9.07 to the laptop by the USB cable
- Select the item (M9.07) which appears on the navigation area on the "FLS
Calibration System" software
- Confirm FW UPGRADE and select the Updated Software
14
Setpoint
Hysteresis
Time
Hysteresis
Setpoint
Time
Setpoint
Hysteresis
Hysteresis
Setpoint
Time
MAX MODE (icon reports MAX)
Flow
Output relaxed
Output energized
PULSE MODE (icon reports PLS)
Pulse duration
Pulse duration
Volume
Volume
Setpoint
Hysteresis
Time
Pulse duration
Volume
Output relaxed
Output energized