FIP FLS M9.07 Instruction Manual

Dual-parameter conductivity and flow monitor & transmitter

Advertisement

Quick Links

FLS M9.07
DUAL-PARAMETER
CONDUCTIVITY AND
FLOW MONITOR & TRANSMITTER
SAFETY INSTRUCTIONS
General Statements
• Do not install and service the product without following the Instruction
Manual.
• This item is designed to be connected to other instruments which can be
hazardous if used improperly. Read and follow all associated instrument
manuals before using with it.
• Product installation and wiring connections should only be performed by
qualified staff.
• Do not modify product construction.
Installation and Commissioning Statements
• Remove power to the instrument before wiring input and output connections.
• Do not exceed maximum specifications using the instrument.
• To clean the unit, use only chemical compatible products.
PACKING LIST
Please verify that the product is complete and without any damage.
The following items must be included:
• M9.07 Dual-Parameter Conductivity and Flow Monitor & Transmitter
• Instruction Manual for M9.07 Dual-Parameter Conductivity and Flow Monitor
& Transmitter
1

Advertisement

Table of Contents
loading

Summary of Contents for FIP FLS M9.07

  • Page 1 FLS M9.07 DUAL-PARAMETER CONDUCTIVITY AND FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements • Do not install and service the product without following the Instruction Manual. • This item is designed to be connected to other instruments which can be hazardous if used improperly.
  • Page 2: Technical Data

    DESCRIPTION The new FLS M9.07 is a dual monitor which combines conductivity and flow measurements. A 4” wide full graphic display shows measured values clearly together with many other useful information. Moreover, due to a multicolor display plus a powerful backlight, measurement status can be determined easily from afar also.
  • Page 3 Electrical • Supply Voltage: 12 to 24 VDC ± 10% regulated • Maximum current consumption: 300 mA • FLS hall effect flow Sensor power: - 5 VDC @ < 20 mA - optically isolated from current loop - short circuit protected •...
  • Page 4: Panel Mounting

    DIMENSIONS PANEL MOUNTING WALL MOUNTING INSTALLATION Mechanical installation The M9.07 Dual-Parameter Conductivity and Flow Monitor & Transmitter is available just in one packaging for panel or wall installation. The panel version is installed using the panel mounting kit (M9.SN1), while the wall mounting version is got fixing the panel mounting version on the wall mounting kit (M9.KWX).
  • Page 5 Panel installation Fix instrument on the panel rotating by hand the fixing snails (M9.SN1). Wall installation Use the panel mounting kit (M9.SN1) to fix the M9.07 on the dedicated frontal cutout of the wall mounting kit (M9.KWX). Tighten front screws of box and waterproof connectors of cables, internally mount caps on screw sites to get a IP65 watertight installation.
  • Page 6 WIRING General recommendation Always ensure the power supply is switched off before working on the device. Make wiring connections according to wiring diagrams. • Terminals accept 26 to 12 AWG (0.08 to 2.5 mm2) • Strip around 10 mm (0.4”) of insulation from the wire tips and tin bare ends to avoid fraying.
  • Page 7: Probe Wiring Diagram

    Refer to dedicated flow sensor manual for its wiring. In general conductivity sensor wires can be connected independently without caring about +IN/REF connectors. In case a temperature sensor (Pt100-Pt1000) is not available, place a brigde connection between 28 - 29 and between 29 - 30. POWER/LOOP WIRING DIAGRAM Stand-alone application, Connection to a PLC with built-in...
  • Page 8 Pt100 - Pt1000 Pt100 - Pt1000 two wires connection three wires connection PT100/1000 PT100/1000 Pt100 - Pt1000 no connection PT100/1000 USB PORT A USB port (type B) is available on the M9.07 PCB. The USB connection allows the updating of device software. To update the software you need: USB cable (M9.KUSB), the interface software "FLS Calibration System”...
  • Page 9: Operational Overview

    Connection to an User User N.O. User AC or DC N.O. N.O. AC or DC AC or DC AC or DC Power Power Power Power User User Imax = 50mA Imax = 50mA Imax = 50mA Imax = 50 lmax = 50mA lmax = 50mA The alarm is off during normal operation and goes ON according to...
  • Page 10 MENU DIRECTORY VIEW LEVEL **** Settings conductivity - temperature Output Settings *** conductivity - flow Flow settings conductivity - temperature - flow item code - software release Calibration conductivity or temperature or flow - analog output 1 conductivity or temperature or flow - analog output 2 Flow Calibration conductivity - direct access to...
  • Page 11 MENU LEVEL Settings Conductivity Unit Temperature Compensation Mode Cell Constant Reference Temperature TDS Factor Temperature Compensation Factor Temperature Unit Manual Temperature Flow Settings K Factor Volume Unit Flow Unit Calibration Temperature Probe Calibration Conductivity Probe Calibration Flow Calibration Auto Calibration Correction Factor Outputs 1 SSR...
  • Page 12: Output Mode

    EDIT LEVEL PUSH BUTTON to modify an item to scroll right to return to Menu without saving to save new settings OUTPUT MODE The M9.07 Dual-Parameter Conductivity and Flow Monitor & Transmitter features 2 solid state relays and 2 mechanical relays in addition to 2 analog output 4-20mA freely.Only the second mechanical relay can be set as a conductivity alarm (icon is 4ALR) related to the feedback of external device managing.
  • Page 13 Digital outputs can be set in the following way: CONDUCTIVITY ON-OFF MODE (icon reports O-F) ON-OFF MODE (icon reports O-F) LOW LEVEL HIGH LEVEL Cond. Cond. Hysteresis Setpoint Hysteresis Output relaxed Time Output relaxed Time Output energized Output energized PROPORTIONAL MODE (icon PROPORTIONAL MODE (icon reports PRP) LOW LEVEL reports PRP) HIGH LEVEL...
  • Page 14: Software Updating

    MAX MODE (icon reports MAX) Cond Setpoint Hysteresis Output relaxed Time Output energized FLOW MIN MODE (icon reports MIN) MAX MODE (icon reports MAX) Flow Flow Hysteresis Setpoint Setpoint Hysteresis Output relaxed Time Time Output relaxed Output energized Output energized WINDOW MODE (icon reports WDW) PULSE MODE (icon reports PLS) Flow...
  • Page 15: Ordering Data

    NOTE: At the end of the procedure restart the instruments in order to refresh M9.07 software (It takes 90 seconds to refresh the SW. Please do not interrupt the restarting process). TO UPDATE NEW UNITS - Download the interface software “FLS Calibration System” and the updated software on www.flsnet.it.
  • Page 16 FIP - Formatura Iniezione Polimeri S.p.A. Loc. Pian di Parata 16015 Casella Genova - Italy Tel. +39 010 96211 Fax +39 010 9621209 www.flsnet.it...