Dual-parameter analog monitor & transmitter (21 pages)
Summary of Contents for FIP FLS M9.07
Page 1
FLS M9.07 DUAL-PARAMETER CONDUCTIVITY AND FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements • Do not install and service the product without following the Instruction Manual. • This item is designed to be connected to other instruments which can be hazardous if used improperly.
DESCRIPTION The new FLS M9.07 is a dual monitor which combines conductivity and flow measurements. A 4” wide full graphic display shows measured values clearly together with many other useful information. Moreover, due to a multicolor display plus a powerful backlight, measurement status can be determined easily from afar also.
Page 3
Electrical • Supply Voltage: 12 to 24 VDC ± 10% regulated • Maximum current consumption: 300 mA • FLS hall effect flow Sensor power: - 5 VDC @ < 20 mA - optically isolated from current loop - short circuit protected •...
DIMENSIONS PANEL MOUNTING WALL MOUNTING INSTALLATION Mechanical installation The M9.07 Dual-Parameter Conductivity and Flow Monitor & Transmitter is available just in one packaging for panel or wall installation. The panel version is installed using the panel mounting kit (M9.SN1), while the wall mounting version is got fixing the panel mounting version on the wall mounting kit (M9.KWX).
Page 5
Panel installation Fix instrument on the panel rotating by hand the fixing snails (M9.SN1). Wall installation Use the panel mounting kit (M9.SN1) to fix the M9.07 on the dedicated frontal cutout of the wall mounting kit (M9.KWX). Tighten front screws of box and waterproof connectors of cables, internally mount caps on screw sites to get a IP65 watertight installation.
Page 6
WIRING General recommendation Always ensure the power supply is switched off before working on the device. Make wiring connections according to wiring diagrams. • Terminals accept 26 to 12 AWG (0.08 to 2.5 mm2) • Strip around 10 mm (0.4”) of insulation from the wire tips and tin bare ends to avoid fraying.
Refer to dedicated flow sensor manual for its wiring. In general conductivity sensor wires can be connected independently without caring about +IN/REF connectors. In case a temperature sensor (Pt100-Pt1000) is not available, place a brigde connection between 28 - 29 and between 29 - 30. POWER/LOOP WIRING DIAGRAM Stand-alone application, Connection to a PLC with built-in...
Page 8
Pt100 - Pt1000 Pt100 - Pt1000 two wires connection three wires connection PT100/1000 PT100/1000 Pt100 - Pt1000 no connection PT100/1000 USB PORT A USB port (type B) is available on the M9.07 PCB. The USB connection allows the updating of device software. To update the software you need: USB cable (M9.KUSB), the interface software "FLS Calibration System”...
Connection to an User User N.O. User AC or DC N.O. N.O. AC or DC AC or DC AC or DC Power Power Power Power User User Imax = 50mA Imax = 50mA Imax = 50mA Imax = 50 lmax = 50mA lmax = 50mA The alarm is off during normal operation and goes ON according to...
Page 10
MENU DIRECTORY VIEW LEVEL **** Settings conductivity - temperature Output Settings *** conductivity - flow Flow settings conductivity - temperature - flow item code - software release Calibration conductivity or temperature or flow - analog output 1 conductivity or temperature or flow - analog output 2 Flow Calibration conductivity - direct access to...
Page 11
MENU LEVEL Settings Conductivity Unit Temperature Compensation Mode Cell Constant Reference Temperature TDS Factor Temperature Compensation Factor Temperature Unit Manual Temperature Flow Settings K Factor Volume Unit Flow Unit Calibration Temperature Probe Calibration Conductivity Probe Calibration Flow Calibration Auto Calibration Correction Factor Outputs 1 SSR...
EDIT LEVEL PUSH BUTTON to modify an item to scroll right to return to Menu without saving to save new settings OUTPUT MODE The M9.07 Dual-Parameter Conductivity and Flow Monitor & Transmitter features 2 solid state relays and 2 mechanical relays in addition to 2 analog output 4-20mA freely.Only the second mechanical relay can be set as a conductivity alarm (icon is 4ALR) related to the feedback of external device managing.
Page 13
Digital outputs can be set in the following way: CONDUCTIVITY ON-OFF MODE (icon reports O-F) ON-OFF MODE (icon reports O-F) LOW LEVEL HIGH LEVEL Cond. Cond. Hysteresis Setpoint Hysteresis Output relaxed Time Output relaxed Time Output energized Output energized PROPORTIONAL MODE (icon PROPORTIONAL MODE (icon reports PRP) LOW LEVEL reports PRP) HIGH LEVEL...
NOTE: At the end of the procedure restart the instruments in order to refresh M9.07 software (It takes 90 seconds to refresh the SW. Please do not interrupt the restarting process). TO UPDATE NEW UNITS - Download the interface software “FLS Calibration System” and the updated software on www.flsnet.it.