2. Electromagnetic compatibility (EMC)

Medical electrical equipment is subject to particular precautions in regard to EMC and must be installed and put into
operation in accordance with the EMC notes included.
W&H guarantees the compliance of the device with the EMC requirements only when used with original W&H accessories
and spare parts. The use of accessories and spare parts not approved by W&H can lead to an increased emission of
electromagnetic interference or to a reduced resistance against electromagnetic interference.
HF communication equipment
Portable HF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to the medical device. Otherwise, degradation of the performance of this medical
device could result.
The medical device may be interfered by other equipment, even if these other devices comply with CISPR (International
special committee on radio interference) emission requirements.
Use of this medical device adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this medical device and the other equipment should be observed to verify that
they are operating normally.
The medical device is not intended for use in the vicinity of HF surgical devices.