Greenlee DM-860A Instruction Manual page 18

Hide thumbs Also See for DM-860A:
Table of Contents

Advertisement

Accuracy
Refer to the "Specifications" section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount) at 23 °C ± 5 °C
(73.4 °F ± 9 °F), 0% to 75% relative humidity. Specifications are for 50,000 counts mode.
True RMS Readings: Voltage and current accuracies are specified from 5% to 100% of the range
unless otherwise specified. Frequency must be within the specified bandwidth for non-sinusoidal
waveforms. Crest factors are as follows:
Crest factor < 2.1:1 at full scale
Crest factor < 4.2:1 at half scale
AC
AC, DC
, and AC + DC
Range
500.00 μA
5000.0 μA
50.000 mA ± (0.5% + 0.05 mA) ± (0.7% + 0.05 mA) ± (2.0% + 0.05 mA) ± (5.0% + 0.05 mA)
500.00 mA ± (0.5% + 0.5 mA)
5.0000 A
10.000 A*
*
10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off).
Range
500.00 μA
5000.0 μA
50.000 mA
500.00 mA
5.0000 A
10.000 A
DC Current
Range
500.00 μA
5000.0 μA
50.000 mA
500.00 mA
5.0000 A
10.000 A*
*
10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off).
18
1.800.561.8187
AC
Current
Accuracy at DC
and 50 Hz to 60 Hz
± (0.5% + 0.5 μA)
± (0.7% + 0.5 μA)
± (0.5% + 5.0 μA)
± (0.7% + 5.0 μA)
± (0.7% + 0.5 mA)
± (0.5% + 0.005 A) ± (0.7% + 0.005 A)
± (0.5% + 0.05 A)
± (0.7% + 0.05 A)
Burden Voltage
(all frequency ranges)
0.15 mV/μA
3.3 mV/mA
45 mV/A
Accuracy
± (0.15% + 0.2 μA)
± (0.1% + 2.0 μA)
± (0.15% + 0.02 mA)
± (0.15% + 0.3 mA)
± (0.5% + 0.002 A)
± (0.5% + 0.02 A)
www.
Accuracy at
Accuracy at
40 Hz to 1 kHz
1 kHz to 20 kHz
± (2.0% + 0.5 μA)
± (2.0% + 5.0 μA)
± (2.0% + 0.5 mA)
Unspecified
Unspecified
Burden Voltage
0.15 mV/μA
3.3 mV/mA
45 mV/A
.com
Accuracy at
20 kHz to 100 kHz
± (5.0% + 0.5 μA)
± (5.0% + 5.0 μA)
± (5.0% + 0.5 mA)
Unspecified
Unspecified
information@itm.com

Advertisement

Table of Contents
loading

Table of Contents