Summary of Contents for Tektronix TTR500 Series

  • Page 1 TTR500 Series Vector Network Analyzers and TCAL500 Calibration Kit Declassification and Security Instructions www.tek.com 077-1256-01...
  • Page 2: Contacting Tektronix

    Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 3 Table of Contents Preface ......................Memory devices ....................Clear or sanitize a non-functioning instrument..............TTR500 Series and TCAL500 Declassification and Security Instructions...
  • Page 4 Table of Contents TTR500 Series and TCAL500 Declassification and Security Instructions...
  • Page 5: Required Documents

    Preface This document helps customers with data security concerns to sanitize or remove memory devices from the Tektronix TTR503A and TTR506A vector network analyzers (VNAs) and the TCAL500 electrical calibration kits. The VNAs have a data storage (memory) device. The instructions in this manual describe how to clear or sanitize the memory device.
  • Page 6 Directly. Describes how data is modified. It means that the user can modify the data. Indirectly. Describes how data is modified. It means that the instrument system resources modify the data and that the user cannot modify the data. TTR500 Series and TCAL500 Declassification and Security Instructions...
  • Page 7: Volatile Memory Devices

    To clear To sanitize SRAM 512 KB Controller Device VectorVu-PC VectorVu-PC U16 on Remove USB Remove USB program switch application application internal cable cable memory, command, commands commands registers transfer TCAL500 calibration memory TTR500 Series and TCAL500 Declassification and Security Instructions...
  • Page 8: Data Export Devices

    128MB data Only not user command Factory modifiable calibration table. Data export devices There are no media or export devices that require declassification or security procedures with this product. TTR500 Series and TCAL500 Declassification and Security Instructions...
  • Page 9 Clear or sanitize a non-functioning instrument Clear or sanitize a non-functioning instrument If your instrument is not functioning, return the instrument to Tektronix for repair. A Tektronix service professional can reprogram the USB device embedded in the VNA. No user data is stored on the instrument when it is disconnected from a power supply.