HeartWare® Ventricular Assist System Patient Manual 
^ Immediate action required, and then call your physician, nurse or VAD coordinator
ALL high alarms will display a message on the Controller Display and the Alarm Indicator will flash 
red. The VAD will stop if the driveline is disconnected or if the controller fails. For a "VAD Stopped" 
alarm, the text message will tell you whether to connect the driveline or change the controller, as 
both of these situations may trigger this alarm.  
The "Controller Failed" alarm indicates a potential controller failure; the controller should be 
exchanged with the back‐up controller. 
WARNING! ALWAYS switch to the backup controller if there is a "Controller Failed" 
alarm. 
The "Critical Battery" alarm is displayed when both batteries only have a few minutes of battery 
time remaining to power your pump. Replace the depleted batteries with fully charged batteries or 
use your AC adapter or DC adapter.  
7.3 Medium Alarms 
The medium alarm starts at a low volume and gets louder over the next minute, unless the Alarm 
Mute button is pressed. A medium alarm is indicated by a flashing YELLOW Alarm Indicator, and the 
text message tells you to call medical personnel. Please call your doctor or nurse immediately to 
receive instructions.   
Alarm  
(Line 1 on controller) 
High Watts  
Electrical Fault 
Low Flow 
Suction 
Controller Fault^ 
Controller Fault^ 
^ Controller Fault indicates a possible controller malfunction, call your clinician for appropriate action. The controller 
may need to be replaced with the back‐up controller.  
* Call your doctor, nurse or VAD coordinator immediately. 
When a medium alarm resolves there is no alarm sound or light displayed in the Alarm Indicator. 
However, the message on the Controller Display will remain until you clear this message by pressing 
the Scroll button  . A new alarm will also clear a resolved medium alarm from the Controller 
Display. 
   
 
Action  
(Line 2 on 
Meaning 
controller) 
A change in the 
Call* 
status of your 
VAD is detected 
Call* 
Possible 
controller 
Call: ALARMS 
malfunction 
OFF* 
Page 28 
Alarm 
Indicator
Alarm Sound 
 
 
Gradual increase in 
volume over the first 
minute if alarm not 
muted.  
Alarm gets louder 
Flashing 
after 5 minutes if 
YELLOW 
alarm not muted. 
Able to mute alarm 
for 5 minutes by 
pressing Alarm Mute 
Button. 
 
 
 
 
  