QUICK REFERENCE GUIDE FOR ALARMS 
When an alarm occurs, two lines of text appear in the Controller Display. The first line tells you 
what the alarm is, and the second line tells you what to do. The chart below shows all potential 
alarms you may see on your controller. 
Alarm Type 
"No Power" Alarm 
High ‐ Critical 
(Flashing Red) 
Medium 
(Flashing Yellow) 
Low 
(Solid Yellow) 
 
HeartWare® Ventricular Assist System Patient Manual 
Alarm Display  (line 1) 
<no message> 
When both power sources (2 batteries or 1 battery and an AC 
adapter or DC adapter) are removed, NO message will display on 
the controller. The "No Power" alarm will sound but the Alarm 
Indicator on the controller WILL NOT light. This indicates your pump 
has stopped. You should immediately connect two power sources. 
VAD Stopped 
VAD Stopped 
Critical Battery 1 
Critical Battery 2 
Controller Failed 
Controller Fault 
Controller Fault 
High Watts 
Electrical Fault 
Low Flow 
Suction 
Low Battery 1 
Low Battery 2 
Power Disconnect 
Power Disconnect 
Action  (line 2) 
<no message> 
Connect Driveline 
Change Controller 
Replace Battery 1 
Replace Battery 2 
Change Controller 
Call 
Call: ALARMS OFF 
Call 
Call 
Call 
Call 
Replace Battery 1 
Replace Battery 2 
Reconnect Power 1 
Reconnect Power 2 