HeartWare Ventricular Assist System Patient Manual page 24

Hide thumbs Also See for Ventricular Assist System:
Table of Contents

Advertisement

HeartWare® Ventricular Assist System Patient Manual 
4. Pull back the white driveline cover from the 
original controller's silver connector. 
 
5. Disconnect the driveline from the original controller by pulling the silver connector away from 
the controller. Do not disconnect by pulling on the driveline cable. A "VAD Stopped" alarm may 
activate. Don't panic. You can silence the alarm after you get your pump restarted. Restarting 
your pump is the priority. 
 
6. Connect the driveline to the new controller (align 
the two red marks and push together). If the "VAD 
Stopped" alarm was active on the new controller, it 
will now resolve. 
 The pump should restart. Verify that the pump 
is working. The RPM, L/min and Watts numbers 
should show on controller display. If your pump 
does not restart, call for medical assistance 
immediately. 
7. To prevent the controller alarm from sounding after 
the power is removed, follow these instructions: 
 If a red alarm adapter is available, insert it into 
the blue connector on the original controller. 
 If no alarm adapter is available: 
o Press and hold the Alarm Mute and Scroll 
buttons on the original controller until a 
"beep" is heard, or for at least 5 seconds. 
o Release the Alarm Mute and Scroll buttons. 
   
 
 
Press and Hold the Alarm Mute and 
 
 
Page 18 
 
 
 
 
Alarm 
 
Scroll Buttons 
 