Immunity test
IEC 60601-1-2
Surge
±1 kV
IEC 61000-4-5
differential
mode
±2 kV
common
mode
Voltage dips,
<5% U
short
(>95% dip in
interruptions,
U
and voltage
cycle
variations on
40% U
power supply
(60% dip in
input lines
U
IEC 61000-4-11
cycles
<5% U
(>95% dip in
U
U
is the A/C mains voltage prior to application of the test level.
T
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150kHz to
80MHz
Radiated RF
3 Vrms
IEC 61000-4-3
80MHz to
2.5GHz
USUM005 Ver. 07
Compliance
Test level
Level
±1 kV
differential
mode
±2 kV
common
mode
<5% U
T
(>95% dip in
U
) for 0.5
) for 0.5
T
T
cycle
40% U
T
(60% dip in
U
) for 5
) for 5
T
T
cycles
<5% U
T
(>95% dip in
U
) for 5 sec
) for 5 sec
T
T
3 A/m
[V1] V
[E1] V/m
UroShield® User Manual
Electromagnetic Environment –
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
T
that of a typical commercial or
hospital environment. The use
of the UroShield® during
continued operation will not be
T
affected by power mains
interruptions, since the
UroShield® is battery powered.
During system charging, it is
recommended that the
T
UroShield® be powered by an
uninterruptible power supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the UroShield®, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
3
5 .
d
V
1
3
5 .
d
E
1
7
d
E
1
Guidance
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
P
27