Icon on Flashing Display
6.5

Disconnecting the Driver from the Actuator

1.
Switch OFF the UroShield® (to prevent the driver from sounding an
alarm when it is disconnected from the actuator).
2.
Hold the driver cable socket in one hand and the actuator cable plug
in the other hand, then pull them apart.
Never attempt to disconnect the driver from the actuator by holding
CAUTION
the driver, actuator, or either of the cables, and pulling on them.
USUM005 Ver. 07
Explanation
This icon appears for
several seconds when
the battery becomes
less than 5% charged.
The UroShield® then
shuts down
automatically.
Actuator malfunction.
The actuator is not
connected to the driver
or is damaged.
UroShield® User Manual
Action Required
Connect the driver to the
mains with the supplied
charger and then switch
ON the UroShield®.
Make sure the actuator
is properly connected to
the driver.
If the actuator is
properly connected and
the malfunction icon is
displayed, replace the
actuator with a new one.