UroShield® User Manual
8
Product Care
8.1
Storage
Before use, store the UroShield® in its shipment box under the following
conditions:
Temp: 10 °C to 30 °C (50 °F to 86 °F)
Humidity: 20% to 55%
The driver battery shelf life is 2 years.
The actuator shelf life is 5 years.
8.2
Environmental Operating Conditions
The recommended environmental operating conditions are:
Temp: 10 °C to 27 °C (50 °F to 80 °F)
Humidity: 40% to 60%
8.3
The Driver
The life expectancy of the driver is 5 years.
To clean the driver use wipes with 70% alcohol or another disinfecting solution.
Do not use solvents (such as acetone) as they may damage the product.
The driver is flame resistant according to UL-94HB. It does not contain toxic,
hazardous, or flammable materials and it will not accelerate a fire.
Nevertheless, the driver is not intended for use in the presence of flammable
liquids.
The driver does not contain recyclable material.
The driver has a 1 year warranty. A faulty unit which is within the warranty
period can be sent to NanoVibronix for replacement.
USUM005 Ver. 07
24
Need help?