Riken Keiki RX-515 Operating Manual

Riken Keiki RX-515 Operating Manual

Portable combination gas detector

Advertisement

PT0E-1015
Portable Combination Gas Detector
RX-515
Operating Manual
MARINE USE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RX-515 and is the answer not in the manual?

Questions and answers

Summary of Contents for Riken Keiki RX-515

  • Page 1 PT0E-1015 Portable Combination Gas Detector RX-515 Operating Manual MARINE USE...
  • Page 2 Carbon dioxide(CO2) :0-20.0vol% This instruction manual is a guide book for operation to use and operate model RX-515. It is kindly requested to read and understand this content by experienced users as well as beginners. The following headline shall be shown to carry out the safety and effective work in this instruction manual.
  • Page 3: Table Of Contents

    C O N T E N T S 1.Caution in operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3 2.Name of each part・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4 3.Operation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7 3-1.Check item before operation・・・・・・・・・・・・・・・・・・・・・ 7 3-2.Preparation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7 3-3.Start-up and end-up・・・・・・・・・・・・・・・・・・・・・・・・・・ 8 3-4.Flow monitor check・・・・・・・・・・・・・・・・・・・・・・・・・・ 9 3-5.Zero point adjustment・・・・・・・・・・・・・・・・・・・・・・・・・ 9 3-6.Measurement・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13 3-7.Peak value・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・...
  • Page 4: 1.Caution In Operation

    1.Caution in operation (For your safety) To maintain the function for safety, follow the following instruction. Warning ・Do not modify nor change the circuit and structure etc. When modify or change, it could not maintain the function. ・This is explosion-proof product. Battery has to be replaced in non-hazardous zone. The battery replacement in hazardous zone shall be out of the scope for explosion-proof product.
  • Page 5: 2.Name Of Each Part

    2 Name of each part . ⑥ ③ ⑤ ② ④ ① GAS OUT GAS IN ⑦ ①POWER/ENTER SW This is used for power ON/OFF and entering the figure. ②PUMP/- SW This is used when make the pump ON/OFF and deverse the figure. ③ZERO/+SW This is used when make zero adjustment and increase the figure.
  • Page 6 【Standard accessories】 ①Carrying case with shoulder strap ②Filter tube with flow monitor and junction tube ③Gas sampling probe ④Gas sampling hose ⑤CO2 scrubber ⑥Alkaline...
  • Page 7 【Optional accessories】 ①Gas sampling hose 30m with plummet ②Gas sampling hose 30m with float type sampling head ③Gas sampling bag 1L for LEL and VOL and calibration gas cylinder ⑤Absorbent cotton ⑥Water proof filter for filter tube (10pcs.) ⑦Steel storage box ⑧Data logging software ⑨Communication cable...
  • Page 8: 3.Operation

    3.Operation 3-1.Check item before operation RX-515 instrument: ・Check that there is no damage for display unit etc. Filter tube with flow monitor: ・Check the dirt of inside cotton filter. When it gets dirt, replace the cotton filter with new one.
  • Page 9: 3-3.Start-Up And End-Up

    3-3.Start-up and end-up When press power switch for 3 sec, the power gets on and this shall start up through the voltage check, filter connection check and sensor connection check. Press power switch for 3 sec Date/time display Stand-by PREPARATION Accessory connection check PROBE OK?...
  • Page 10: 3-4.Flow Monitor Check

    3-4.Flow monitor check When the power gets ON, the pump shall start operation check that suction works properly by filter tube with flow monitor. Float At normal sucking(Can see float) Not sucking(Can not see float) 3-5.Zero point calibration Zero point calibration for this instrument shall be done for CH4(Methane), O2(Oxygen), CO(Carbon monoxide) and that of CO2 carbon dioxide by separating them.
  • Page 11 <FLOW CHART OF ZERO POINT ADJ> Display ・Press ZERO key for 3 sec. ZERO key HOLD KEY ・When press ENTER key, CH4, O2, CO zero cal FRESH AIR? starts ESC key Move to zero adj ENTER key for CO2 adj ESC ADJ ZERO ・Calibration 【Zero set OK】...
  • Page 12 《Caution at use of SCRUBBER (CO removed filter)》 ・ Mount CO2 scrubber to the sampling side of filter tube with flow monitor. ・ When operate, put erect the CO2 scrubber. When suck the sample gas horizontally, CO2 in air may not be sucked. ×...
  • Page 13 ・ Do not let high density CO2 sucked into instrument. ・ Do not breathe in from air during zero calibration. ・ After use of CO2 scrubber, shut off the ventilation from atmosphere. When the absorbent agent is mixed with air, the absorption capability will be dropped by CO2 in air. ・...
  • Page 14: 3-6.Measurement

    3-6.Measurement Approach the sampling hose to the sampling site for measurement and leave sampling gas sucked. ・Take care that any twist on the sampling hose can not be made. ・Stop sampling hose at the point of measurement. ・From the hose length consider the delay time and read out the reading after the reading gets stable.
  • Page 15: 3-7.Peak Value

    3-7.Peak value When press PEAK key during operation, the peak valve from power on to the current point (CH4, CO, CO2 = MAX, O2=MIN) shall be shown. Display PEAK value display PEAK key P E A K When hold and press ESC key for 3 sec during the peak value display, the peak value shall be cleared.
  • Page 16: 3-9.Operation Finished

    Operation display ・Press pump key for 3 sec. PUMP key HOLD KEY ・Pump work will be off. PUMP OFF PUMP key ・Press pump key for 3 sec. HOLD KEY PUMP ON ・Pump work will be on. Hold and press for 5 sec and back to the operation display.
  • Page 17: 4.Alarm

    4.Alarm 4-1.Alarm display ①At 1 alarm, the figure will blink and “Warring” will blink. ②At 2 alarm, the figure will blink and “Alarm” will blink. ③At over alarm, the figure will blink and “Over” will blink. 《1 alarm》 Alarm≧1 alarm set(CH 、CO、CO )...
  • Page 18: 4-3.Trouble Display

    4-3.Trouble display (1)Cow flow ・When low flow trouble triggers, “COW FLOW” is displayed and pump stops working. ・Buzzer sounds continuously. ・After checking the cause, reset low flow alarm by “ESC” key. The pump will re-start and return to the operation display. Pump ON Pump OFF LOW FLOW...
  • Page 19 (3)System trouble ・When the program is troubled, “SYSTEM FAIL” will be shown. ・Buzzer will sound continuously. ・After “SYSTEM FAIL” is shown, it does not accept other than “Power off”. SYSTEM FAIL S A V E Power off available only by power key (4)Sensor trouble ・When sensor is troubled, “FAIL SENSOR”...
  • Page 20: 5.Each Function

    5.Each function 5-1.Setting mode Operation display ・Press MODE key for 3 sec. MODE key HOLD KEY Displayed in the following order ① 1 alarm ①1st alarm display ② 2 alarm MODE key ③ Date display ④ Data logger remaining W A R N I N G -...
  • Page 21: 5-1-1.Alarm Setting

    5-1-1.Alarm setting alarm display W A R N I N G ・Input the password【3002】. ・When press『+』or『-』, the number will be changed. PEAK key ・Fix the number by ENTER key and move to next digit. ・After fixing the last digit, move to alarm setting (When make the password error, it will show “ERROR”.
  • Page 22: 5-1-2.Date/Time Setting

    5 -1-2.Date/time setting Date/time display D A T E / T I M E ・First, set the year. PEAK key ・The set number shall blink. ・The change of number is set by 『+』 or『-』key. ・ Press ENTER key and the number is D A T E /...
  • Page 23: 5-3.Data Logger

    5-3.Data logger Data logger function has following 5 kinds. (1)Interval trend The change of measuring gas shall be recorded until the power on into off. Recording time can 100 times data. When it is over 100 times, the old data shall be deleted and record the new data by up-dating.
  • Page 24: 6.Maintenance Check

    6.Maintenance check For this correct operation, follow to the following procedure. As this is safety instrument. Be sure to make the regular maintenance. If any trouble should take place, contact RIKEN KEIKI or nearest agent. 6-1.Battery replacement Warning ・ For battery replacement, make it non-hazardous area by all means after turning off the power.
  • Page 25: 6-2.Gas Calibration (Span Adjustment)

    6-2.Gas calibration (Span adjustment) ・ Hold and pressing ZERO and PUMP key simittaneously in the operation mode. ZERO key PUMP key ・First, CH4 display shall be shown. HOLD KEY ・ When press ENTER key, “CH4” gas calibration shall start. ・When press『+』key, it shall change to ESC key gas calibration of 「...
  • Page 26 Danger ・Conpressed cylinder gas is used for gas calibration and take care at handling. ・At calibration by combustible gas, do not approach the fire. ・Exhant the remaining gas in the sampling gas appropriately. Caution Before gas calibration, be sure to make zero adjustment. 《Handling of sampling bag and canned gas》...
  • Page 27: 6-3.Filter Replacement

    (1) It can not calibrate. (2) Reading can not go back if make air calibration. (3) The reading is unstable. When the sensor replacement is needed, contact RIKEN KEIKI or nearest agent. 6-5.Daily check and regular check (1)Daily ・Is there any damage for switches, lamps, display and panel ? ・Pump suction performance check (Does the pump moving sound correctly ? Can any abnormal...
  • Page 28: 6-6.The Replacement Of Parts

    6-6.The replacement of parts The following parts have a life and each, regular replacement shall be recommended. When replace, contact RIKEN KEIKI or nearest agent. 《Recommended consumable parts list for replacement》 Name Check Replacement frequency frequency (Year) Sensor pack 6 months...
  • Page 29: 6-7.Storage Or Treatment Not To Use For A Long Period

    When dispose the waste, follow the regional low etc locally and do it in the same category wite the industrial waste through the appropriate procedure any harmful materials are not used except oxygen sensor and carbon monoxide sensor. For O2 sensor and CO sensor waste, contact RIKEN KEIKI or nearest agent.
  • Page 30: 8.Trouble Shooting

    ・Re-start pump operation. Press PUMP SW until LCD working. will display “PUMP ON”. ・Pump connection fails. ・Check connection condition and contact RIKEN KEIKI for repair. Gas is not sucked ・Filter or hose etc are removed ・Check connection status of filter, hose and filth or clogged.
  • Page 31: 9.Definition Of Words

    9.Definition of words Vol% Gas density is expressed by the unit of a part. Gas density is expressed by the unit of a part per. 10,000ppm=1vol% %LEL Suppose to put the Lower Explosive Limit as 100 and at that time, combustible gas density is expressed by the unit of a part per hundred.
  • Page 32: 10.Specifications

    10.Specifications Refer to the specifications for separate catalog overleaf. Model RX-515 Gas detected Methane(CH4) Oxygen(O2) Carbon monoxide(CO) Carbon dioxide(CO2) Detection principle NDIR(Non-Dispersive Galvanic cell Electro-chemical NDIR(Non-Dispersive infrared) cell infrared) 0-20.0vol% Detection range 0-100%LEL(0.5%LEL) 0-2vol%(0.01vol%) 0-25vol%(0.1vol%) 0-1000ppm(1ppm) (digit) 0-100vol%(0.5vol%) -5vol%(0.05vol%) -20vol%(0.1vol%) Operating Temp.
  • Page 33: 11.Detection Principle

    11.Detection principle NDIR method (Non-Dispersive Infrared) 11-1 . The model is based on NDIR method (Non-Dispersive Infrared) and this structure is shown below. The infrared beam emitted from the light source passes through the measuring cell, and optical band pass filter which can pass the absorption wave of measuring gas and attains to the infrared sensor.
  • Page 34: 11-3.Electro-Chemical Cell Method

    11-3.ELECTRO-CHEMICAL CELL method This gas detector applies an electrochemical sensor. The sample gas is electrolyzed by the electrolyzed cell added with specific electric potential (bias voltage) and detected from the electrolyzed current generated at that time. The electrochemical sensor is designed to keep the interface between electrode and electrolyte at a constant potential (Bias voltage) and is the method to electrolyze gas directly.