HomeFill® System
Pulse Setting
Pressures
Inlet Pressure Range
Max Fill Pressure
Internal Regulated Pressure
Safety Burst Disk
20
1, 1.5, 2, 2.5, 3, 4, 5, 6
Conserving (Savings) Ratio,
Bolus Size and Accuracy
@ 20 BPM
Equivalent Flow Rate
1 L/min setting:
1.5 L/min setting:
2 L/min setting:
2.5 L/min setting:
3 L/min setting:
4 L/min setting:
5 L/min setting:
6 L/min setting:
Sensitivity
0.130 to 0.150 in (0.330 to 0.381 cm) H
Delivery
"ON" time
"OFF" time
2068 - 14479 kPa (300 - 2100 psig)
14479 kPa (2100 psig)
345 kPa (50 psig)
Bursts at 20.8 — 23.2 Mpa at 22°C (3025–3360 psig at 72°F)
Pulse Volumes (L)
Minimum
0.014
0.021
0.029
0.036
0.043
0.056
0.071
0.084
< 125 ms
< 300 ms
Nominal
Maximum
0.0165
0.0248
0.0330
0.0413
0.0495
0.0660
0.0825
0.0990
O
2
0.019
0.029
0.038
0.047
0.057
0.076
0.095
0.114