Cylinder Operation - Invacare HomeFill HF2ECE9 User Manual

Oxygen conserving cylinders
Hide thumbs Also See for HomeFill HF2ECE9:
Table of Contents

Advertisement

Available languages

Available languages

HomeFill® System
3.
With the other hand, grasp the outer ring (sleeve) of the
connector fillport and push DOWN.
4.
Lift up on the cylinder/regulator assem-
bly to remove from the connector fillport.
5.
When the cylinder fillport is disconnected from the connector
fillport, release the outer ring (sleeve) of the connector fillport
and use two hands to remove the cylinder from the compressor
cradle.
WARNING!
– The fillport cover on the connector fillport and
the cylinder fillport MUST be replaced after
filling and whenever not in use. If either of the
fillport covers are missing contact your healthcare
provider before using.
6.
Place the fillport covers onto the connector fillport and cylinder
fillport.
7.
Examine the cylinder gauge to ensure cylinder is full (needle
pointing into GREEN area). If the cylinder is not full, perform
the following:
14
Repeat STEPS 1–9 of Connecting the Cylinder to the
Compressor, page 11.
If cylinder is not full after second attempt to fill, contact
the provider.
8.
Slowly close the shutoff valve fully (clockwise).

4.2 Cylinder Operation

WARNING!
– NEVER smoke in an area where oxygen is being
administered.
– NEVER use near any type of flame or
flammable/explosive substances, vapours or
atmosphere.
1.
Prior to each use, inspect product for visible damage. Refer to
3.4 External Examination, page 9 . DO NOT use if any damage
is found.
1176148-A

Advertisement

Table of Contents
loading

This manual is also suitable for:

Homefill hf2ece6

Table of Contents