Summary of Contents for Physio Control LIFEPAK CR2
Page 1
(PLM) System. Approvals can be obtained from the PLM system which displays the selected approvers, their approval roles, and approval dates. PROPRIETARY AND CONFIDENTIAL DOCUMENT FOR PHYSIO-CONTROL, INC. USE ONLY. This document is property of Physio-Control and may not be used, reproduced, published or disclosed to others without authorization from Physio-Control. Owner...
Serial Label Information ............................... 10 Symbols ..................................10 General Danger and Warnings Statements ......................... 12 LIFEPAK CR2 Defibrillator Description.......................... 15 LIFEPAK CR2 Defibrillator Specifications ........................15 CR2 Device Description ..............................20 Physical Description and Features ..........................20 Functional Descriptions..............................21 Modes of Operation ..............................
Page 3
Applying Updates with USB connection ........................54 Monitor Update Status ..............................59 View Software Version ..............................61 Viewing LIFEPAK CR2 Log Files ............................63 View Log from LIFENET System ............................ 63 View Log from LIFELINK ............................... 65 View WCT Log ................................68 Post AED Event Services ..............................
Page 4
4 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Replacement Procedure ..............................85 Supply and Replacement Parts ............................ 85 Lid Replacement ................................86 Handle or Carrying Case Replacement......................... 87 USB Cover Replacement .............................. 88 Battery Replacement ..............................
Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Preface This Service manual describes how to support the following tasks for the LIFEPAK CR2 defibrillator: • Connecting to LIFELINK or LIFENET System • Confirming connectivity • Checking self-test results •...
Page 6
LIFENET Device Agent, Operating Instructions PCO file Physio Case Object file, a data format for a combination of device data and user-entered data. Data in this format can be imported and exported by Physio-Control data management applications. SmartDesk Data Solutions Technical Support Wireless Configuration Tool is a software Application to configure Wi-Fi connection in the LIFEPAK CR2®...
800 442 1142 https://lifelink.lifenetsystems.com Service Personnel Qualifications Service technicians must be properly qualified and thoroughly familiar with the operation of the LIFEPAK CR2 defibrillator. They must meet at least one of the following requirements (or the equivalent): • Physio-Control Service Depot technicians •...
99512-0000XX* NOTE: * XX in the PN and REF can be different on each unit depending on the order configuration of the device. Warranty To obtain a detailed warranty statement, contact your local Physio-Control representative or go to www.physio- control.com Tamper Proof Label If the tamper-proof label shown below is broken or missing, the warranty may be void.
For reference, the general danger and warning statements from the LIFEPAK CR2 Defibrillator Operating Instructions, 3321515. d) A brief description of the LIFEPAK CR2 Defibrillator and its functions, and the device’s significant physical and performance characteristics. e) Information concerning circuit diagrams, component lists available for authorized service representatives...
Page 11
11 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State SYMBOL DESCRIPTION Consult instructions for use Warning, high voltage General warning Single use only Not made with natural rubber latex Use By date shown: YYYY-MM-DD (on electrodes) Expiration date: YYYY-MM-DD (on battery) Do not recharge battery Lithium manganese dioxide battery...
Serial number Part number General Danger and Warnings Statements For reference, the following are the general danger and warning statements found in the LIFEPAK CR2 Defibrillator Operating Instructions, 3321515. Explosion Hazard Do not use this defibrillator in the presence of flammable gases or anesthetics.
Page 13
Do not rapidly key EMS radios on and off. Refer to Separation Distances (on page Error! Bookmark not defined.) for recommended distances of equipment. Contact Physio-Control Technical Support if assistance is required. Possible Electrical Interference This defibrillator should not be used adjacent to or stacked with other equipment.
Page 14
14 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Possible Electrical Interference Using cables, electrodes, or accessories not specified for use with this defibrillator may result in increased emissions or immunity from electromagnetic or radio frequency interference (RFI) which could affect the performance of this defibrillator or of equipment in close proximity.
LIFEPAK CR2 Defibrillator Description The LIFEPAK CR2 defibrillator is designed for indoor and outdoor use in a stationary environment. Fully automatic and semi-automatic models are available. After the electrode pads are applied, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any responder assistance. The semi- automatic model evaluates the heart rhythm but requires the responder to press the shock button if a shockable rhythm is detected.
Page 16
16 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Defibrillator Waveform Biphasic Truncated Exponential with voltage and duration compensation for patient impedance. Patient Impedance 10 to 300 ohms. If the detected impedance is outside this range, the Range user is prompted to check the electrode pads and connector.
Page 17
17 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Biphasic waveform at 200 joules, nominal PATIENT I1 (A) I2 (A) I3 (A) I4 (A) (MS) (MS) IMPEDANCE (Ω) 58.1 21.0 20.9 10.6 30.0 14.8 14.8...
Page 18
18 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Shock Ready Time Note: These times apply for a device that has just reached a low battery state (Readiness indicator has stopped flashing). Fully automatic device: The time to charge to 360 joules after electrode pads are applied is 35 seconds or less.
Page 19
19 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Battery Operating 0° to 50°C (32° to 122°F) Temperature Long-term Storage 0° to 25°C (32° to 77°F) Temperature Short-term Storage -30° to 60°C (-22° to 140°F) for up to 1 week Temperature Relative Humidity 5 to 95% non-condensing...
20 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State CR2 Device Description Physical Description and Features Four Control 7000545 Revision A...
State Date 11/14/2017 State Functional Descriptions The LIFEPAK CR2 defibrillator is a portable, battery-powered therapeutic device. It consists of the following components: • One PCB – the printed circuit board controls defibrillator functions, including ECG signal processing, mode control, rhythm and motion analysis, energy storage capacitor charging, impedance sensing, voice prompting, generation of biphasic defibrillation waveform, and patient event recording.
22 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Modes of Operation Operation Description Open the Lid AED turns on and starts talking. LANGUAGE Allows user to change the voice prompts to different language if a dual-language is Button installed on the device.
4. Each battery maintains a real-time clock in UTC-0 time. All real time events on the CR2 will be converted to local time from LIFENET System webpages, log file, event file and reports based on the device’s location. 5. There is no battery status indicator on the LIFEPAK CR2 device. See Battery Maintenance for more information.
5. If patient care is transferred to emergency medical personnel, electrode pads can be transferred to another LIFEPAK defibrillator (except LIFEPAK 500). Connectivity Module The LIFEPAK CR2 defibrillator has a built-in connectivity module that can connect to LIFELINK or LIFENET System using Wi-Fi, or Cellular, or USB connection. •...
25 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State If this number is present, the device has Wi-Fi If this additional pair of ID and IC numbers is present, the device has cellular capability and the SIM card is activated.
Page 26
11/14/2017 State The LIFENET ® System is a cloud-based platform designed by Physio-Control for Pre-Hospital and Hospital. It provides a complete software infrastructure that supports all LIFEPAK devices and 3 party products to: • Share critical patient data and reduce time-to-treatment for STEMI patients.
Page 27
11/14/2017 State Wireless Configuration Tool The Wireless Configuration Tool is a computer application created by Physio-Control to help customers store Wi-Fi settings on the CR2 device so that the device can communicate with either LIFENET or LIFELINK. LIFENET Device Agent LIFENET®...
Page 28
28 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State DT EXPRESS™ Data Transfer Software DT Express Data Transfer Software is used to download patients’ cases from LIFEPAK devices to PCO data format. CODE-STAT™...
Page 29
29 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State AED Event Viewer AED Event Viewer is a Mobile/PC Application to view AED event data from a specific incident, route data to CODE- STAT, and trigger LIFENET Alert Client Software.
30 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Initial Device Setup Overview The following diagram describes the initial device setup steps performing by the customers upon receiving the CR2. The printed literatures within the shipping box are designed to help the customers to perform device setup to bring the device to Ready state.
CR2 Trainer Device Description CR2 Trainer Device Overview The LIFEPAK CR2 Trainer, AED Training System, provides realistic training in the use of the LIFEPAK CR2 defibrillator without the actual charge and discharge of electrical energy. The trainer is recommended so users can practice using the LIFEPAK CR2 defibrillator in simulated cardiac arrest situations.
LIFEPAK CR2 Trainer will have a one year limited warranty. • Future software updates will be orderable from Physio-Control pre-loaded on a USB device. The customer can use the USB device to reprogram trainers. Each new software update will have a unique part number for order.
33 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 3323900-080 21250-000007 ASSY, LID, LOCALIZED, TRAINER, LPCR2, FRENCH 3323900-100 21250-000008 ASSY, LID, LOCALIZED, TRAINER, LPCR2, DUTCH 3323900-120 21250-000009 ASSY,LID,LOCALIZED,TRAINER,LPCR2,ES-SPAIN-LATAM 3323900-180 21250-000010 ASSY, LID, LOCALIZED, TRAINER, LPCR2, SWEDISH 3323900-200 21250-000011 ASSY, LID, LOCALIZED, TRAINER, LPCR2, DANISH...
NOTE: CR2 supports 2.4 GHz Wi-Fi frequency. It does not support 5 GHz frequency. Download and Install Wi-Fi Configuration Tool Overview: Wi-Fi Configuration Tool is a PC application made by Physio-Control to help customers connect Physio- Control devices to the Wi-Fi network so that the devices can communicate with LIFENET system. This procedure describes how to download and install Wi-Fi Configuration Tool.
35 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 5. Follow the onscreen instructions to finish installing Wi-Fi Configuration Tool. 6. If an older version is detected during the installation, remove the previous version and repeat from step 3. Store Wi-Fi Settings on CR2 Overview: The CR2 has no screen and keyboard, to setup Wi-Fi configuration on CR2, the Wi-Fi Configuration Tool software application is used from a computer, and a USB cable is used to transfer the Wi-Fi settings to the CR2.
Page 36
36 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State This Wi-Fi Configuration Tool is intended to reach LIFENET product servers. When using this tool to test connectivity in LIFELINK or LIFENET test server, you must enter the test server URL in the tool. To do so, in the country field, select United States and then press Ctrl + F10.
Page 37
37 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. Get your Proxy Server information from IT and enter into the following page so that the Wi-Fi Configuration Tool can reach the LIFELINK or LIFENET website. NOTE: The Wi-Fi Configuration Tool will only store Proxy Server information on the computer;...
Page 38
38 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 5. Connect the USB cable between the CR2 and computer. 7000545 Revision A...
Page 39
39 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 6. Open the CR2 lid. 7. Follow the onscreen instructions from the Wi-Fi Configuration Tool. 8. After selecting a Wi-Fi network name and entering Security key, click Test Connection to test if the CR2 can communicate with the Wi-Fi network hotspot and reach LIFENET System.
Page 40
40 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 9. If the Connection Test Failed screen appears, follow the 5 common solutions on the screen to resolve the issue. CR2 MAC address Note: If customer’s IT needs to know the MAC address of CR2, use a wrong password for the Wi-Fi network and let Test Connection failed, the CR2 MAC address will be displayed in Connection Test Failed window.
Page 41
41 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 7000545 Revision A...
Page 42
42 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 11. Click Next when Update successful screen appears. 12. For a site has multiple CR2s that need to connect to LIFENET, select Configure Another Device. Otherwise, select Close the Lid, disconnect the USB cable, and Close Wi-Fi Configuration Tool.
43 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Note: A multiple Wi-Fi networks can be stored in CR2 by repeating above procedure. The CR2 will be automatically detect and connect to whichever Wi-Fi network is in the range. Confirm Connection to LIFELINK or LIFENET Overview: After the Wi-Fi settings have been stored in the CR2, perform this procedure at the intended CR2 location to make the CR2 check in to LIFELINK or LIFENET.
44 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 2. Open the lid. 3. Wait until the voice prompts start and then press and hold the LANGUAGE and CHILD MODE buttons simultaneously. 4.
Page 45
45 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State • From LIFELINK login: Go to the Equipment page and verify that the CR2 status is “Ready” in green. • From LIFENET login: Go to the Devices page and verify that the CR2 status is “Ready” in green. 7000545 Revision A...
46 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Monthly Auto-Check-In Overview: This session explains how often and when the CR2 will send device data to LIFELINK or LIFENET System to update device status. The CR2 will send data when any of the following events occur: 1.
47 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. If the CR2 loses connectivity and misses the monthly Auto-Check-In to LIFELINK or LIFENET System, the Device Readiness status will be set to Outdated. To fix the Outdated status, refer to Confirm Connection to LIFELINK or LIFENET to perform manual check-in.
Page 48
48 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 7000545 Revision A...
Page 49
49 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 5. To remove Wi-Fi settings from the CR2, select More Options > Erase all Wi-Fi settings from the device > Next. The Wi-Fi settings on the CR2 will be erased and the Update Successful screen will appear. 6.
Change Setup Options or Update Software Overview: All LIFEPAK CR2 devices are shipped with customer-defined settings, e.g., language, therapy energy level, CPR Settings, and other options. If a customer wants to change the factory default settings or update the software, the changes must be made through LIFELINK or LIFENET System.
Page 51
51 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. Click on icon to open Software and Setup Options page 5. Select a profile (settings and software version are site-specific) 6. In the profile page, click on Edit Setup Options 7000545 Revision A...
Page 52
52 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 7. Then make changes to the setup options, and click on Save 8. To print or keep a record of site setup options, click on 9.
Page 53
53 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 10. To see which devices are linked to this profile, click the 'View Devices' icon, Steps (using LIFENET login as an example) 1. Login to LIFENET (both customers’ accounts or service accounts can perform this function) 2.
54 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Applying Updates with Wi-Fi or Cellular Connection Updates are not applied to CR2 until CR2 check-in to LIFELINK or LIFENET Systems. CR2 with access to a wireless connection or cellular network will check-in to LIFELINK or LIFENET automatically once a month.
Page 55
55 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Steps: 1. If LIFENET Device Agent has been download to the computer before, skip to Step 4 2. To download LIFENET Device Agent, login to LIFELINK -> Resource Center -> Connection Tools -> LIFENET Device Agent ->...
Page 56
56 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State NOTE: When using this tool to test LDA in LIFELINK or LIFENET test server, you must enter the test server URL in the tool. To do so, in the country field, select United States and then press Ctrl + F10. This should pop-up an input box where you can type in the test server URL.
Page 57
Released State Date 11/14/2017 State 6. Click Next in the Device List window ® 7. Select LIFEPAK CR2 Defibrillator and click Next, 8. Connect the USB cable from the CR2 to the computer, then open device lid 7000545 Revision A...
Page 58
58 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 9. If it is first time using LDA with CR2 USB cable, the USB driver software will be automatically installed to the computer. Then trying to establish the connection between LDA and CR2 via USB cable. During this time, following screen appears.
59 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 11. When download is complete, the following screen appears. Monitor Update Status Overview: All CR2 devices of the same model will have the same profile (software version and setup options) to ensure consistent device behavior across one site (the settings are site-specific;...
Page 60
60 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Monitor Update Status from LIFELINK: 1. Login to LIFELINK website 2. Click on Sites from left menu 3. Highlight a site and click to open the Site view 4.
61 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Monitor Update Status from LIFENET: To check if update has been applied to the device, from Assets -> from View Assets by dropdown menu -> select Device software configuration ->...
Page 62
62 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State View device software version from LIFENET System 1. login to LIFELINK System 2. click Assets 3. Select View Assets by Device software configuration 4.
State Date 11/14/2017 State Viewing LIFEPAK CR2 Log Files Overview: To view CR2 device log files, service personnel will need to have LIFENET or LIFELINK accounts. View Log from LIFENET System From My Network menu, click Devices. In the Actions column, select the device row and click to open the View Device Log page.
Page 64
64 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State There are four types of Device Logs for viewing, depending on the LIFENET System user role. Not all logs are available for viewing in all roles. For the LIFENET Account Support User role, all logs are available for viewing. 1.
View Log from LIFELINK View Device Log: For connected LIFEPAK CR2 devices, the Device Log can be viewed from the LIFELINK website. 1. Test Log – provides information about internal tests that are automatically run on the device including test type, test results, and fault codes.
Page 66
66 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Device Log 7000545 Revision A...
Page 67
67 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State View AED Usage Log: The AED Usage Log shows data such as the device model and serial number, event date, and the date post event reports were sent to the Post Event Report Recipient (if applicable).
68 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. The AED Event Log Item window shows the Event detail for date, time, SN, and Site Name. View WCT Log Overview: When the CR2 cannot connect to LIFELINK of LIFENET, there is a communication log in the WCT which will provide service more information for troubleshooting connection errors.
Page 69
69 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. When the Connect to Wi-Fi Network screen appears, select Advanced Setup, and Next 7000545 Revision A...
Page 70
70 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 5. Select Read LIFEPAK CR2 diagnostic data, and Next to see Communication Details 7000545 Revision A...
71 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 6. If customer needs support with connection issues, instruct the customer to copy the log entries into a text file: • selecting the first line in the log window, •...
Page 72
72 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State NOTE: 1. The “AED in Use” notification email will only be sent to Program Coordinator who is assigned in LIFELINK. 2. The “AED Post Event” email will be sent to the Medical Direcor with two reports attached in the email: CPR report and Event Log Report.
Page 73
73 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 3. CPR Report Example: 7000545 Revision A...
74 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 4. Event Log Example: 5. If the customer wants to see continuous wave form summary (ECG data), follow Case 3 below to download Post Event data from the CR2 to the computer.
Case 3 (non-connected CR2 or USB-only CR2): For non-connected LIFEPAK CR2 devices, including USB-only, if customers purchase AED Post-Event Service, an FSR will be sent on-site to use DT EXPRESS to manually download the AED Event from the CR2. The DT EXPRESS version which supports the CR2 is 6.1 or greater.
Page 76
76 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Prerequisites: • Computer with internet connection • DT EXPRESS 6.1 or greater installed on computer • CR2 that has been used in a patient event Required Tools: •...
Page 77
77 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 2. Click QUIK-STEP. 7000545 Revision A...
Page 78
78 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 3. Select LIFEPAK™ CR2 AED from Download Wizard. 4. Connect USB cable from CR2 to computer. 7000545 Revision A...
Page 79
79 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 5. Open CR2 lid and wait for download to complete. 7000545 Revision A...
Page 80
80 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 6. Select desired files to download, and click Next>. NOTE: The default setting of DT EXPRESS is set to only show unsent cases. To see the cases that have been sent to LIFENET system from the computer, contact Smartdesk to get a batch file, LPCR2DownloadUnsentFilesOnlyUI_Disable.bat, to modify the registry key.
11/14/2017 State 7. The Downloaded PCO files will be stored in this location: C:\ProgramData\Physio-Control\Resuscitation Suite\10\Download\. The PCO files can be manually imported to CODE-STAT for viewing. 8. If a viewer prefers to have the case in PDF format, select Export to export the case to PDF files. The export folder is pre-configured in Configuration Properties.
82 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State NOTE: • Self-test pass: 3 audible tones sound and the green Readiness indicator is flashing every 6 seconds through a small hole located on top of the lid, the self-test has been completed and passed. •...
- Ped test Battery Maintenance Overview: There is no battery status indicator on the LIFEPAK CR2 device. The battery status is monitored through a combination of self-tests, automatic email notifications to the owner with battery status for connected devices, and voice prompt diagnostics.
Page 84
84 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State • Battery Status: For devices that are connected to LIFENET System, the battery status can be viewed on the LIFENET website in the Device view with LIFENET service role. •...
Overview: The Performance Inspection Procedure (PIP) is a set of manual and computer assist test procedures. The PIP is used for an operational, closed-case evaluation of the LIFEPAK CR2 defibrillator to determine if the defibrillator is operating within the required specifications. The PIP will be performed in a regularly scheduled preventive maintenance routine.
86 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Catalog Number Description 3323374-000 11141-000165 Kit, Replacement Battery LPCR2 3323372 11101-000021 Replacement Electrode Kit (Leonhard Lang) 3323375-000 11512-000001 Kit, Replacement, Lid, LPCR2 3323376-000 11512-000002 Kit, Replacement, Handle, LPCR2 3323377-000...
87 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Handle or Carrying Case Replacement To replace the Handle or Carrying Case: 1. Remove three screws from the handle. 2. Replace the handle with a new handle or with a carrying case. 3.
88 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State USB Cover Replacement Install USB Cover onto Rear Case by inserting prong into small hole and push it all the way in. Battery Replacement To replace the Battery: 1.
89 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Refer to OI, “Replacing the Electrodes” for the proper steps to replace the electrode tray. NOTE: When the electrodes are used or expired, the Readiness indicator will not flash until the electrode tray is replaced. When a new electrode tray is installed, and after closing the lid, the Readiness indicator will flashes every 6 seconds.
90 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Start CR2 Readiness light is not flashing Open the Lid Can you hear voice prompt? Replace Battery Open the Lid Can you hear voice prompt? Press Two buttons Language + Child...
6. Check if a COM port is automatically assigned to the CR2 7. If no COM port is assigned to the CR2, reboot the computer and repeat the steps. When LIFEPAK CR2 USB Port is visible, proceed to Wi-Fi Configuration ToolWi-Fi Configuration Tool or LIFENET Device Agent tasks.
92 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 2. If Wi-Fi connection from the smart phone is working, check that the Wi-Fi settings on the CR2 are correctly entered. Refer to Store Wi-Fi Settings on CR2 or Change or Remove Wi-Fi Settings on the CR2 for details.
Page 93
93 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State After completion of "Wi-Fi Connection Initiated", and attempt to establish a Wi-Fi connection to Connection In Progress LIFENET. During connection attempt, device beeps every 5 seconds. If Wi-Fi connection cannot be established from Unable to Establish Wi-Fi CR2 to Wi-Fi access point.
94 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State An error has occurred while attempting to connect LIFENET Connection Error to LIFELINK or LIFENET System. Contact SmartDesk for support. Once LIFENET connection has been established. First HTTP response is an OK.
Page 95
95 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 0x40040 AOKLOGGEREVENT_AdcCalibrationFail 0x40041 AOKLOGGEREVENT_WakeUpAlarm7DayTimeout 0x41700 SHOCKEVENT_OverVolt 0x41701 SHOCKEVENT_DumpTmrExp 0x41705 SHOCKEVENT_ChargeCmpltVcapChkFail 0x41706 SHOCKEVENT_EnergyCalibrationCrcFailed 0x41707 SHOCKEVENT_EnergyCalibCalculationFailed 0x41708 SHOCKEVENT_ChargeTimedOut 0x42108 PREPROCEVENT_EcgCalibrationCrcFailed 0x42109 PREPROCEVENT_DeltaZCalibrationCrcFailed 0x4210A PREPROCEVENT_ImpedanceCalibrationCrcFailed 0x4210B PREPROCEVENT_EcgCalibCalculationFailed 0x4210C PREPROCEVENT_DeltaZCalibCalculationFailed...
Page 96
96 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 0x4252A STEVENT_IsoRelayCloseFail 0x4252C STEVENT_IsoRelayOpenFail 0x42536 STEVENT_HBridgePhs1CondFail 0x42538 STEVENT_HBridgePhs2CondFail 0x4253A STEVENT_ChargeRateFail 0x4253C STEVENT_VCapSysAokMeasureFail 0x4253E STEVENT_VCapFlbckAokMeasureFail 0x42540 STEVENT_RamCompleteFail 0x42542 STEVENT_SpeakerFail 0x42544 STEVENT_SwLidFail 0x42546 STEVENT_SwOnOffFail 0x42548 STEVENT_SwShockFail 0x4254A STEVENT_SwLangFail...
97 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State Tables of useful information LIFENET System URL This is LIFENET System URL in each country. Use this list for customer’s Proxy server whitelist exemption. Country Country Code...
HIPAA Security Information for the LIFENET® System List of Reference Document and Tool Name Location LIFEPAK CR2 Defibrillator eCentral or http://www.physio-control.com/ Operating Instructions, 3321515 LIFEPAK CR2 Getting Started eCentral or http://www.physio-control.com/ Guide, 3321528 LIFEPAK CR2 Wireless Setup eCentral or http://www.physio-control.com/ Guide, 3322174 LIFENET System Help File (PDF) From website, Click Help >...
99 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State LIFELINKcentral Help (PDF) From website, Click > Contents > LIFELINKcentral Help (PDF), then save it to your computer. • Wi-Fi Configuration Tool Download Wi-Fi Configuration Tool from LIFELINK account, select Resource Center, and then select Wi-Fi Configuration Tool, and click on to download.
100 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State System Context Diagram 7000545 Revision A...
101 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State System Component View 7000545 Revision A...
Page 102
102 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 7000545 Revision A...
Customer Number: ______ Dear [Name]: On [Date], Physio-Control, Inc. has confirmed that this device was not used in accordance with applicable operating instructions or was not used in the intended environment or setting; specifically the device was [enter specific failure circumstance]. As such, Physio-Control is unable to certify the long-term reliability of a device subjected to this condition.
Page 104
104 of Title Service Manual, LPCR2 Page Service Product Drawing 3323299 Type Released State Date 11/14/2017 State 3330033_A: Cust Ltr Warranty Void, CR2, Danish 3330034_A: Customer Letter, Warranty Void, CR2, Chinese 3331868_A: Customer Letter, Warranty Void, LPCR2, Simplified Chinese 3332134_A: Cust Ltr Warranty Void, LPCR2, Arabic 3332135_A: Cust Ltr Warranty Void, LPCR2, Greek 3332136_A: Cust Ltr Warranty Void, LPCR2, Hebrew 3333060_A: Cust Ltr Warranty Void, LPCR2, Hungarian...