Assessing The Results - Elcometer AVU User Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
4 PERFORMING THE TEST (continued)
5
Apply pressure until the required test pressure is reached on the adhesion gauge and compare the
adhesion gauge's reading with the reading on the Elcometer AVU display - see Section 5 "Assessing
the Results" on page 5.
6
To switch between measurement units, press the "MPa/psi" softkey.
The Elcometer AVU will provide adhesion readings for 20mm dollies irrespective of which dolly adaptor is
fitted - see below for additional information.
ELCOMETER 506 DOLLY ADAPTOR FOR 50MM ACTUATOR SKIRT (Part Number: T99923936)
When using the Elcometer 506 Analogue Adhesion Tester with 50mm dolly adaptor, the reading on the
Elcometer AVU display should be compared with the reading on the 20mm scale of the Elcometer 506
analogue pressure gauge.
When using the Elcometer 506 Digital Adhesion Tester with 50mm dolly adaptor, the dolly size should be
set to 20mm before test - refer to the Elcometer 506 User Guide for instructions on how to select the
20mm dolly size.
ELCOMETER 106 SCALE 6 DOLLY ADAPTOR (Part Number: T99923938)
When testing the Elcometer 106 Pull-Off Adhesion Tester with the Scale 6 dolly adaptor, the Elcometer
AVU reading should be multiplied by a conversion factor of 6.4516 before comparing with the adhesion
gauge reading.
For other manufacturers' gauges, please refer to the manufacturers' instructions for adhesion readings
using 20mm dollies.
www.elcometer.com
R
4

Advertisement

Table of Contents
loading

Table of Contents