4.2 Use with a 8921 conductimeter

The use of a type 8398 inductive probe connected to a type 8921 2-channel
conductimeter requires that this latter be fitted with a software version greater than or
equal to the following versions:
Software
Standard
Concentration
Special
To install your new software version, proceed as follows:
x Remove the transmitter's front panel (4 screws).
x On the rear of the front panel, there are two circuit boards. Remove the first of
these two circuit boards (4 screws).
x At the rear of this printed circuit board, carefully change the integrated circuit
(EPROM) represented on the figure opposite.
x Check that the memory has been inserted in the correct position, otherwise it
would be irreversibly damaged.
x Re-fit the front panel and initialise the instrument's default parameters using
command 900.
x Reprogram the instrument to your configuration and configure command C101
(102) with Argument 2 , which corresponds to the 8398 probe's cell constant,
equal to 2.35.
EPROM
Minimum
version no.
3.76
1.04
4.07
Inductive Probe 8398
Memory reference
08921=A=6200
08921=A=6400
08921=A=6305