Service and maintenance
Cleaning
The flowmeter's access door seal is removable for quick and easy cleaning.
Spraying the load cell directly with high pressure may damage the load cell.
Spare parts
The following are the recommended spare parts. As each may be available in more than one
option, consult Siemens Milltronics for assistance in ordering spare parts. In all correspondence
please quote the serial number of the unit for which the parts are required.
● Load cell
● Sensing plate
Replacing a load cell
Remove the load cell/sensing plate assembly from the housing for replacement or maintenance:
1. Remove the access cover and secure access door.
2. Dis-connect the load cell.
3. Hold the load cell by the upper cross piece.
4. Remove load cell mounting screws (2 places), and then lift load cell/sensing plate assembly
out of the housing. The cable will need to be pulled through the conduit (if installed) and any
cable glands that prevent removing the assembly.
32